Quantifying the influence of bone density and thickness on resonance frequency analysis: an in vitro study of biomechanical test materials.

Thibaut Bardyn1, Philippe Gédet, Wock Hallermann, Philippe Büchler.   

Abstract

PURPOSE: Resonance frequency analysis (RFA) offers the opportunity to monitor the osseointegration of an implant in a simple, noninvasive way. A better comprehension of the relationship between RFA and parameters related to bone quality would therefore help clinicians improve diagnoses. In this study, a bone analog made from polyurethane foam was used to isolate the influences of bone density and cortical thickness in RFA.
MATERIALS AND METHODS: Straumann standard implants were inserted in polyurethane foam blocks, and primary implant stability was measured with RFA. The blocks were composed of two superimposed layers with different densities. The top layer was dense to mimic cortical bone, whereas the bottom layer had a lower density to represent trabecular bone. Different densities for both layers and different thicknesses for the simulated cortical layer were tested, resulting in eight different block combinations. RFA was compared with two other mechanical evaluations of primary stability: removal torque and axial loading response.
RESULTS: The primary stability measured with RFA did not correlate with the two other methods, but there was a significant correlation between removal torque and the axial loading response (P < .005). Statistical analysis revealed that each method was sensitive to different aspects of bone quality. RFA was the only method able to detect changes in both bone density and cortical thickness. However, changes in trabecular bone density were easier to distinguish with removal torque and axial loading than with RFA.
CONCLUSIONS: This study shows that RFA, removal torque, and axial loading are sensitive to different aspects of the bone-implant interface. This explains the absence of correlation among the methods and proves that no standard procedure exists for the evaluation of primary stability.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20162104

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  7 in total

1.  Evaluation of accuracy in implant site preparation performed in single- or multi-step drilling procedures.

Authors:  Nadine Marheineke; Uta Scherer; Martin Rücker; Constantin von See; Björn Rahlf; Nils-Claudius Gellrich; Marcus Stoetzer
Journal:  Clin Oral Investig       Date:  2017-12-17       Impact factor: 3.573

2.  Influence of cortical bone anchorage on the primary stability of dental implants.

Authors:  Anne Kelly de Oliveira Nicolau Mantovani; Ivete Aparecida de Mattias Sartori; Luciana Reis Azevedo-Alanis; Rodrigo Tiossi; Flávia Noemy Gasparini Kiatake Fontão
Journal:  Oral Maxillofac Surg       Date:  2018-06-06

3.  Biomechanical evaluation of oversized drilling technique on primary implant stability measured by insertion torque and resonance frequency analysis.

Authors:  Gorka Santamaría-Arrieta; Aritza Brizuela-Velasco; Felipe J Fernández-González; David Chávarri-Prado; Yelko Chento-Valiente; Eneko Solaberrieta; Markel Diéguez-Pereira; José-Antonio Vega; Jaime Yurrebaso-Asúa
Journal:  J Clin Exp Dent       Date:  2016-07-01

4.  Retrospective clinical study of ultrawide implants more than 6 mm in diameter.

Authors:  Jeong-Kui Ku; Yang-Jin Yi; Pil-Young Yun; Young-Kyun Kim
Journal:  Maxillofac Plast Reconstr Surg       Date:  2016-08-05

5.  The effects of bone density and crestal cortical bone thickness on micromotion and peri-implant bone strain distribution in an immediately loaded implant: a nonlinear finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2016-06-28       Impact factor: 2.614

6.  Effects of implant tilting and the loading direction on the displacement and micromotion of immediately loaded implants: an in vitro experiment and finite element analysis.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Sadami Tsutsumi; Tadaaki Kirita
Journal:  J Periodontal Implant Sci       Date:  2017-08-28       Impact factor: 2.614

7.  Evaluation of Primary Stability of Cylindrical and Tapered Implants in Different Bone Types by Measuring Implant Displacement: An In vitro Study.

Authors:  Tsutomu Sugiura; Kazuhiko Yamamoto; Satoshi Horita; Kazuhiro Murakami; Tadaaki Kirita
Journal:  Contemp Clin Dent       Date:  2019 Jul-Sep
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.