BACKGROUND: Over the last few years, differential ion mobility spectroscopy (DMS) has become an important tool in medical research. There are attempts to find markers for specific diseases in exhaled air, using this technology as a non-invasive early diagnosis. OBJECTIVE: In the present research, exhaled air from 78 patients with known diagnosis and 39 control persons were tested with a DMS system from Sionex. RESULTS: Bronchial asthma showed a pattern of 6 characteristic points in a discriminant analysis. Patients with diagnosed hypertension showed a characteristic pattern with 4 points, hypothyroidism 2 points; increased LDL cholesterol 3 points, and type II diabetics treated with insulin 4 spots. No significant differences with respect to the control group were found in chronic obstructive pulmonary disease patients. The DMS pattern in the tested asthmatics showed a partial change depending on different medications used. CONCLUSION: Differential ion mobility spectroscopy offers promise as a helpful diagnostic tool.
BACKGROUND: Over the last few years, differential ion mobility spectroscopy (DMS) has become an important tool in medical research. There are attempts to find markers for specific diseases in exhaled air, using this technology as a non-invasive early diagnosis. OBJECTIVE: In the present research, exhaled air from 78 patients with known diagnosis and 39 control persons were tested with a DMS system from Sionex. RESULTS:Bronchial asthma showed a pattern of 6 characteristic points in a discriminant analysis. Patients with diagnosed hypertension showed a characteristic pattern with 4 points, hypothyroidism 2 points; increased LDL cholesterol 3 points, and type II diabetics treated with insulin 4 spots. No significant differences with respect to the control group were found in chronic obstructive pulmonary diseasepatients. The DMS pattern in the tested asthmatics showed a partial change depending on different medications used. CONCLUSION: Differential ion mobility spectroscopy offers promise as a helpful diagnostic tool.
Scientific research at the University of California in Irvine showed that breath analysis is an effective, non-invasive method to determine the blood glucose level. Using an analytical method, primarily developed to measure air contamination, researchers found markedly increased methyl nitrate concentrations in exhaled air from children with type 1 diabetes [1]. They then tested ten children with type 1 diabetes with high as well as declining blood glucose levels after administration of insulin. Exhaled air from these patients was tested for 100 different markers in the ppt-range (parts per trillion). It has been shown that the concentration of methyl nitrate in patients is 10 times higher than in those with the normal glucose concentration and correlated with the level of blood sugar. In follow-up studies, researchers plan to correlate hyperglycemia with other gases in exhaled air to establish a complete profile of exhaled markers in diabetes [1]. These studies seem a prelude to the development of a breath analyzer for diabetics to control their insulin usage.Different metabolized products, e.g., from tumor cells, from thoracic diseases and from different bacteria, but also from changes in metabolism due to medication, lead to different detectable metabolites in exhaled air [2,3]. Today, it is possible to find and measure markers in exhaled air for diabetes [4,5], breast cancer [6], tuberculosis [7], and lung cancer [8,9]. The concentration of most markers is very low, so most diagnostic standard tests do not capture them.Already in 1971, Pauling et al [10] described an investigation of exhaled air using gas chromatography (GC) and found 250 different substances. In today's literature, the most described technology to measure exhaled air is gas chromatography using different detectors, such as mass sensitive detectors (MSD) [11], photo-ionization detectors (PID), flame-ionization detectors (FID), and also ion traps [12]. In most cases, enrichment of the samples is required because the concentration of analytes is only a few parts per million (ppm) or even parts per billion (ppb), which is usually below the detection limit. Enrichment is usually done with chemical or adsorptive binding to specific adsorbent materials with acceptable technical effort. Cooling traps are also used to concentrate the sample; however the technical complexity is high.Other methods to analyze breath are different forms of mass spectrometry, proton-transfer reaction mass spectrometry (PTR-MS), and the selected ion flow tube mass spectrometry (SIFT-MS). Both the methods are useful for on-line analysis, have detection limits in a lower ppb-range, and short detection times. Technology is similar in both methods [13]. In the SIFT-MS, ions are created in a microwave plasma ion source which is external to the flow tube. The ions are then extracted from the ion source, selected according to their mass-to-charge ratio using a quadruple mass filter, injected into a flowing carrier gas sampled by a downstream pinhole orifice, mass analyzed, and counted by a differentially-pumped quadruple mass spectrometer system [14,15].Differential mobility spectroscopy (DMS) is a newer technique. It is capable of detecting metabolites at very low concentrations. Ion mobility spectroscopy detects different substances such as aldehydes, halogens, poly-aromatic substances and nitric compounds [16]. DMS gained importance to detect explosive and toxic agents during the Irak war. Today's uses include environmental monitoring, detection of narcotic or explosive markers in airports, also in the forensic chemical and criminalistic labs. Also, the international space station uses a DMS to monitor the air [17].
Materials and methods
Detection of metabolites in exhaled air requires coupling a differential mass spectrometer with a multicapillary column to reach a pre-separation of the known 250 substances [10]. Furthermore, a heated molecular trap with activated charcoal is used to enrich the sample. Coupled differential mass spectrometers like this allow the identification and quantification of volatile metabolites in exhaled air down to concentrations in the ppb level within 5 min. A DMS from Sionex in Bedford was used for the present study. The principle is shown in Figure 1. Probe collection is by a pump drawing air into the system through a 3-way valve and preconcentration trap. A second pump creates a constant flow of carrier gas through a molecular filter, a carbon filter, and the ionization chamber. In a second step, the column is loaded. The 3-way valve changes its pathway to allow flow of the carrier gas through the concentration trap and the column. At the same time, the trap is heated to the desired temperature to allow a desorption of molecules of interest from the trap to the column. In a last step, the column is heated and molecules flow through the ionization chamber to the DMx sensor depending on their retention time.
Figure 1
Differential mass spectroscopy.
Differential mass spectroscopy.After passing through the 63Ni ionization chamber, the charged ions pass with the carried gas through a tunable electric field and land at a positive or negative electrode as shown in Figure 1. The generated electric signal is transmitted to a PC by an analog-digital converter.Different publications show the existence of various volatile organic compounds (VOC) in expired air at tracer levels. Breath consists of different groups of compounds. One consists of exogenous substances which must be taken into account with the measurement. These analytes are inspired from room air or picked up with food substances or even by dermal absorption of different fragrances or other active components, e.g., from sanitary products. They may be transmitted to exhaled air by blood circulation [18]. Measurements with the DMS can be disturbed by such analytes in high concentration and may mask the wanted marker. For our research, we defined a useful test time after eating. Patients had to wait for 2 h after eating to be included into the study.Our tests were conducted as a controlled, open-labeled monitoring study in 3 test centers in and around Berlin. Voluntary test persons with known diagnosis were selected, questioned, and measured. The patient questionnaire listed medication, eating habits, known diseases, and risk factors plus gender, size, weight, age, barometric pressure, temperature and humidity: see Table 1
Table 1
Questionnaire to access symptomatic and treatment details of volunteers.
1
Initials
2
Gender
3
Age
4
Size
5
Weight
6
Time from last meal
7
Nutrition
8
Risk factors
9
Diagnosis
10
Other diseases
11
Current lung disease
12
Barometric pressure, temperature, humidity
Questionnaire to access symptomatic and treatment details of volunteers.To prevent a direct patient's contact with the DMS, we used a special breathing tube, connected to the DMS by a teflon tube. The test procedure for the patient was a deep inspiration followed by a slow expiration through the sample tube. The internal pump collected 20 ml out of the collection tube. The pump was started after expiration of the ventilatory dead space to make sure that only air from the central and lower airways was taken [19].Analyzing time was 4 min and 30 s. Reference samples were collected from 39 healthy persons. Spectrums were compared and related to the questionnaire; a cluster analysis was carried out. A total of 117 participants were measured, 78 patients and 39 blank samples.For the interpretation, discriminant analysis with previous cluster analysis was used, because the individual group association was known, and only differences between the groups were searched. Not only were the relations between variables studied, but the goal was to predict unknown values of dependant variables by means of declarative variables.
Results and Discussion
Several patients showed different diseases at the same time. This resulted in the detection of 188 diseases in 117 patients. The result is shown in Figure 2. Seventy eight patients included 17 smokers, 47 overweight (BMI > 25), and 9 with adipositas (BMI > 30); 23 male and 55 female patients participated in this pilot study. Among the 39 healthy subjects, there were 20 smokers, 16 overweight, none with heavy adipositas; 11 man and 28 women. The mean age of patients was 59 ± 14 years; the reference group was 41 ± 22 years.
Figure 2
Occurrence of different diseases in the tested population.
Occurrence of different diseases in the tested population.There was no age or weight dependant pattern in the discriminant analysis. Commonly, cholesterol is measured in the lab from blood. Cholesterol comes from food or from the liver which generates cholesterol. Increased cholesterol level itself is not a disease, but is an indicator for increased risk of arteriosclerosis. The cholesterol value also is of interest for increased risk of coronary cardiac disease and circulatory problems. Fourteen out of our 117 patients reported increased LDL cholesterol levels and took medication to lower their cholesterol. These patients showed matching patterns in the cluster analysis. Twenty one of the other tested subjects showed the same pattern. None of these were treated for high cholesterol, nor took any cholesterol lowering medication. This shows that the pattern really corresponded to the high cholesterol level rather than medication.In healthy individuals, the thyroid gland delivers the hormones thyroxin and triiodothyronine into the blood. This requires iodine. The daily need for an adult is approximately 200 μg per day. Sufficient uptake with food is not always warranted in countries like Germany [20]. Eighteen of our patients reported to be treated for thyroid hypofunction. They all showed a specific pattern which is shown in Figure 3 (Group 2). Noticeably, 4 of the 18 patients also had increased LDL cholesterol which is known to go along with thyroid hypofunction.
Figure 3
Cluster analysis. The upper pictures show the positive and negative spectrums from one patient, without allocating the different points. The lower pictures show the same spectrum after allocating several points to groups using cluster analysis.
Cluster analysis. The upper pictures show the positive and negative spectrums from one patient, without allocating the different points. The lower pictures show the same spectrum after allocating several points to groups using cluster analysis.No specific pattern was found in the COPDpatients. Twenty two of these patients had diverse metabolites in their breath. Seven of the COPDpatients reported to have coronary heart disease. They, plus another 11 patients with coronary heart disease, could be grouped into a total of 18 patients with established coronary disease as shown in Figure 3 (Group 5). Six of them also showed thyroid hypofunction.Chronic airway diseases, like bronchial asthma, are serious diseases with rising prevalence. The immune system and the nervous system are involved into the pathomechanisms of these diseases. Both systems interact through different mediators, some of which are the neutrophins. The four neutrophins BDNF, NGF, NT-3, and NT-4/5 play an important role in the activation and differentiation of immune cells and have additional influence on structural cells, like epithelium cells or fibroblasts/myofibroblasts. Asthmatics show an increased concentration of these neutrophins [21]. Our asthmapatients showed an interesting fact: a total of 38 asthmatics had to be assigned to 3 different groups: Group 4, Group 6 (Figure 3), and Group 7 (not shown in Figure 3). A comparison of the patients' data showed a difference to be dependent on the individual's medication. Depending on the kind of medication, only 3 out of the 6 metabolites present in all asthmatic patients could be traced. No subject of the control group and no asthmatic patient, according to the questionnaire, could be related to Groups 4, 6, and 7.VOCs found in connection with diabetes mellitus are described in publications by Amann and Smith [4] and Phillips et al [5]. A very noticeable pattern could be connected to diabetes type IIpatients. However, based on the high amount of non-detection of this pattern and concord with patients from Group 2 taking medications, it can be assumed that the medication is responsible for the pattern. The patients from Group 2 regularly take insulin.During the study time, slight dislocation of the patterns occurred which were most likely created by changes in humidity and barometric pressure. The patterns were slightly offset proportionally. Because using the cluster and discriminant analyses only the patterns were detected, the absolute coordinates could be neglected.
Summary
1/ The DMS technology is today widespread as a detection method for explosives, chemical weapons, and drugs. Space research uses the DMX to monitor the air in the space shuttle. An analysis of exhaled air is already studied in various places.2/ Analysis of exhaled air is already commonplace for O2, CO2, NO, and other gases. The use of a DMS to analyze volatile organic compounds is underway and a user-friendly instrument is the next development goal.3/ Our preliminary results with asthmatics, patients with hypothyreosis, hypertension, diabetes type II, and hypercholesterolemia show that the Sionex unit is useful for breath analysis. The use of this unit to monitor the effects of medication or for the early detection of a diseases, e.g., like lung carcinoma, is possible.
Conflicts of interest
The authors declare that they have no competing interests.