Anesthetic agent vapor analyzers and propellants of pressurized meter-dose inhalers.

Kavaraganahalli Mukundarao Deepak1, Kiran Kini.   

Abstract

Anesthetic agent analyzers fail when drug from a HFA propelled pMDI is administered. Most anesthetic vapor analyzers use infrared absorption at 3.3 micron. At this wavelength, substance like methane has been reported to interfere with the accuracy of measurement of anesthetic vapor. The anesthetic gas monitor 1304 (Brüel and Kjaer) which functions at 10.3-13 micron wavelength was not affected by methane. Is it possible, HFA with its structural similarity to inhaled anesthetic agents may be responsible for faulty reading of anesthetic vapor concentration in two of our monitors? Further evidence is needed to support this finding. Anesthesiologists need to be ever vigilant and recognize the need for smarter designs of anesthetic agent analyzers with changing array of drugs administered during anesthetic management.

Entities:  

Mesh:

Year:  2010        PMID: 20130969     DOI: 10.1007/s10877-010-9220-x

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  5 in total

Review 1.  Withdrawal of albuterol inhalers containing chlorofluorocarbon propellants.

Authors:  Leslie Hendeles; Gene L Colice; Robert J Meyer
Journal:  N Engl J Med       Date:  2007-03-29       Impact factor: 91.245

Review 2.  Pressurised metered dose inhalers versus all other hand-held inhaler devices to deliver beta-2 agonist bronchodilators for non-acute asthma.

Authors:  F S Ram; D M Brocklebank; J White; J P Wright; P W Jones
Journal:  Cochrane Database Syst Rev       Date:  2002

3.  Methane influences infrared technique anesthetic agent monitors.

Authors:  E Mortier; G Rolly; L Versichelen
Journal:  J Clin Monit Comput       Date:  1998-02       Impact factor: 2.502

Review 4.  Inhaled bronchodilator administration during mechanical ventilation.

Authors:  Alexander G Duarte
Journal:  Respir Care       Date:  2004-06       Impact factor: 2.258

Review 5.  Bronchodilator delivery with metered-dose inhaler during mechanical ventilation.

Authors:  D Georgopoulos; E Mouloudi; E Kondili; M Klimathianaki
Journal:  Crit Care       Date:  2000-07-11       Impact factor: 9.097

  5 in total
  1 in total

1.  Interference of anesthesia monitoring equipment by HFA propellants. Letter to the editor.

Authors:  Phillip D Levin; Alexander Avidan
Journal:  J Clin Monit Comput       Date:  2010-06-09       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.