An alternative option for elderly patients with a small aortic annulus: the 16 mm ATS valve.

Yoshitsugu Nakamura1, Kiyoharu Nakano, Osamu Tagusari, Go Kataoka, Yoshimasa Seike, Satoru Domoto, Yujiro Ito.   

Abstract

BACKGROUND AND AIM OF THE STUDY: The optimal procedure and prosthesis remains debatable for aortic valve replacement (AVR) in high-risk elderly patients in whom the aortic annulus is too small to allow a standard AVR procedure with even the smallest sized bioprosthetic valve available. Herein are reported the early and mid-term results of standard AVR using a 16 mm ATS Advanced Performance (AP) mechanical heart valve.
METHODS: The medical records of 10 patients (mean age 75 +/- 5 years; range: 64-79 years) in whom 16 mm ATS AP valves had been implanted in the supra-annular position were reviewed retrospectively. Preoperatively, the mean body surface area was 1.46 +/- 0.1 m2 (range: 1.21-1.69 m2); mean logistic EuroSCORE 16.1 +/- 14.6% (range: 4.32-53.2%); mean peak pressure gradient (peak PG) across the aortic valve 98 +/- 28 mmHg; mean diameter of the aortic annulus 19.3 +/- 1.2 mm; mean fractional shortening 37 +/- 10%; and mean left ventricular myocardial mass index (LVMI) 173 +/- 34 g/m2.
RESULTS: There were no hospital deaths, and one late death. The postoperative course was uneventful in all cases, except for one patient who developed respiratory failure. The NYHA functional class was improved from 3.0 +/- 0.7 before surgery to 1.4 +/- 0.5 postoperatively. A significant decrease in the peak PG was observed postoperatively compared to preoperative data (p < 0.01); mean values at two weeks, and at one and two years after surgery were 35 +/- 12, 35 +/- 10, and 33 +/- 10 mmHg, respectively. A significant decrease in the mean LVMI was also noted postoperatively (p < 0.01); mean values at two weeks, and at one and two years after surgery were 134 +/- 35, 110 +/- 17, and 114 +/- 22 g/m2, respectively. The mean effective orifice area index was 0.79 +/- 0.2 cm2/m2. During the mean follow up period of 56 +/- 23 months, all patients received oral anticoagulation with warfarin, and there were no cases of bleeding or thomboembolic complications.
CONCLUSION: Satisfactory early and mid-term outcomes were noted following AVR with the 16 mm ATS AP valve in high-risk elderly patients with a small aortic annulus.

Entities:  

Mesh:

Year:  2009        PMID: 20099719

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  3 in total

1.  Early and late outcomes of AVR with aortic annular enlargement in octogenarian.

Authors:  Yuki Okamoto; Kazuo Yamamoto; Tsutomu Sugimoto; Shinpei Yoshii
Journal:  Gen Thorac Cardiovasc Surg       Date:  2015-05-14

2.  Cardiac energetics analysis after aortic valve replacement with 16-mm ATS mechanical valve.

Authors:  Tomoki Ushijima; Yoshihisa Tanoue; Takayuki Uchida; Sho Matsuyama; Takashi Matsumoto; Ryuji Tominaga
Journal:  J Artif Organs       Date:  2014-05-25       Impact factor: 1.731

Review 3.  Current status and future perspectives of prosthetic valve selection for aortic valve replacement.

Authors:  Hiroshi Furukawa; Kazuo Tanemoto
Journal:  Gen Thorac Cardiovasc Surg       Date:  2013-05-31
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.