MRI assessment of pacing induced ventricular dyssynchrony in an isolated human heart.

Michael D Eggen1, Michael G Bateman, Christopher D Rolfes, Stephen A Howard, Cory M Swingen, Paul A Iaizzo.   

Abstract

This study demonstrates the capabilities of MRI in the assessment of cardiac pacing induced ventricular dyssynchrony, and the findings support the need for employing more physiological pacing. A human donor heart deemed non-viable for transplantation, was reanimated using an MR compatible, four-chamber working perfusion system. The heart was imaged using a 1.5T MR scanner while being paced from the right ventricular apex (RVA) via an epicardial placed lead. Four-chamber, short-axis, and tagged short-axis cines were acquired in order to track wall motion and intramyocardial strain during pacing. The results of this study revealed that the activation patterns of the left ventricle (LV) during RVA pacing demonstrated intraventricular dyssynchrony; as the left ventricular mechanical activation proceeded from the septum and anterior wall to the lateral wall, with the posterior wall being activated last. As such, the time difference to peak contraction between the septum and lateral wall was approximately 125 msec. Likewise, interventricular dyssynchrony was demonstrated from the four-chamber cine as the time difference between the peak LV and RV free wall motion was 180 msec. With the ongoing development of MR safe and MR compatible pacing systems, we can expect MRI to be added to the list of imaging modalities used to optimize cardiac resynchronization therapy (CRT) and/or alternate site pacing.

Entities:  

Mesh:

Year:  2010        PMID: 20099368      PMCID: PMC2893377          DOI: 10.1002/jmri.22050

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  10 in total

1.  Visualizing myocardial function using HARP MRI.

Authors:  N F Osman; J L Prince
Journal:  Phys Med Biol       Date:  2000-06       Impact factor: 3.609

2.  Isolated four-chamber working swine heart model.

Authors:  E Chinchoy; C L Soule; A J Houlton; W J Gallagher; M A Hjelle; T G Laske; J Morissette; P A Iaizzo
Journal:  Ann Thorac Surg       Date:  2000-11       Impact factor: 4.330

3.  Government viewpoint: U.S. Food & Drug Administration: Pacemakers, ICDs and MRI.

Authors:  Owen P Faris; Mitchell J Shein
Journal:  Pacing Clin Electrophysiol       Date:  2005-04       Impact factor: 1.976

4.  Comparison of myocardial velocities obtained with magnetic resonance phase velocity mapping and tissue Doppler imaging in normal subjects and patients with left ventricular dyssynchrony.

Authors:  Jana G Delfino; Mohit Bhasin; Robert Cole; Robert L Eisner; John Merlino; Angel R Leon; John N Oshinski
Journal:  J Magn Reson Imaging       Date:  2006-08       Impact factor: 4.813

5.  Right ventricular pacing can induce ventricular dyssynchrony in patients with atrial fibrillation after atrioventricular node ablation.

Authors:  Laurens F Tops; Martin J Schalij; Eduard R Holman; Lieselot van Erven; Ernst E van der Wall; Jeroen J Bax
Journal:  J Am Coll Cardiol       Date:  2006-09-27       Impact factor: 24.094

6.  Detrimental ventricular remodeling in patients with congenital complete heart block and chronic right ventricular apical pacing.

Authors:  Jean-Benoît Thambo; Pierre Bordachar; Stephane Garrigue; Stephane Lafitte; Prashanthan Sanders; Sylvain Reuter; Romain Girardot; David Crepin; Patricia Reant; Raymond Roudaut; Pierre Jaïs; Michel Haïssaguerre; Jacques Clementy; Maria Jimenez
Journal:  Circulation       Date:  2004-12-06       Impact factor: 29.690

7.  In vitro studies of human hearts.

Authors:  Alexander J Hill; Timothy G Laske; James A Coles; Daniel C Sigg; Nicholas D Skadsberg; Sarah A Vincent; Charles L Soule; William J Gallagher; Paul A Iaizzo
Journal:  Ann Thorac Surg       Date:  2005-01       Impact factor: 4.330

Review 8.  Cardiac magnetic resonance assessment of mechanical dyssynchrony.

Authors:  Robert H Helm; Albert C Lardo
Journal:  Curr Opin Cardiol       Date:  2008-09       Impact factor: 2.161

9.  Mechanical dyssynchrony or myocardial shortening as MRI predictor of response to biventricular pacing?

Authors:  Iris K Rüssel; Jaco J M Zwanenburg; Tjeerd Germans; J Tim Marcus; Cornelis P Allaart; Carel C de Cock; Marco J W Götte; Albert C van Rossum
Journal:  J Magn Reson Imaging       Date:  2007-12       Impact factor: 4.813

10.  Safety of magnetic resonance imaging of patients with a new Medtronic EnRhythm MRI SureScan pacing system: clinical study design.

Authors:  Richard Sutton; Emanuel Kanal; Bruce L Wilkoff; David Bello; Roger Luechinger; Inge Jenniskens; Michael Hull; Torsten Sommer
Journal:  Trials       Date:  2008-12-02       Impact factor: 2.279

  10 in total
  3 in total

1.  Novel imaging of atrial septal defects in isolated human hearts.

Authors:  Stephen A Howard; Jason L Quill; Michael D Eggen; Cory M Swingen; Paul A Iaizzo
Journal:  J Cardiovasc Transl Res       Date:  2013-02-15       Impact factor: 4.132

Review 2.  Imaging in the context of replacement heart valve development: use of the Visible Heart(®) methodologies.

Authors:  Michael G Bateman; Paul A Iaizzo
Journal:  Cardiovasc Diagn Ther       Date:  2012-09

3.  Cardiovascular magnetic resonance imaging of isolated perfused pig hearts in a 3T clinical MR scanner.

Authors:  Andreas Schuster; Amedeo Chiribiri; Masaki Ishida; Geraint Morton; Matthias Paul; Shazia T Hussain; Boris Bigalke; Divaka Perera; Tobias Schaeffter; Eike Nagel
Journal:  Interv Med Appl Sci       Date:  2012-12-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.