| Literature DB >> 20098540 |
Rajesh A Kinhikar1, Zubin Master, Dipak S Dhote, Deepak D Deshpande.
Abstract
A Helical Tomotherapy (HT) Hi-Art II (TomoTherapy, Inc., Madison, WI, USA) has been one of the important innovations to help deliver IMRT with image guidance. On-board, mega voltage computed tomography (MVCT) detectors are used for imaging and dosimetric purpose. The two objectives of this study are: (i) To estimate the dosimetric and general capability (TomoImage registration, reconstruction, contrast and spatial resolution, artifacts-free image and dose in TomoImage) of on-board MVCT detectors. (ii) To measure the dosimetric parameters (output and energy) following major repair. The MVCT detectors also estimated the rotational output constancy well. During this study, dosimetric tests were repeated after replacing MVCT detectors and the target. fixed-gantry/fixed-couch measurements were measured daily to investigate; the system stability. Thermoluminescense dosimeter (TLD) was used during both the measurements subsequently. The MVCT image quality with old and new detectors was comparable and hence acceptable clinically. The spatial resolution was optimal and the dose during TomoImage was 2 cGy (well within the manufacturer tolerance of 4 cGy). The results of lateral beam profiles showed an excellent agreement between the two normalized plots. The output from the rotational procedure revealed 99.7% while the energy was consistent over a period of twelve months. The Hi-Art II system has maintained its calibration to within +/- 2% and energy to within +/- 1.5% over the initial twelve-month period. Based on the periodic measurements for rotational output and consistency in the lateral beam profile shape, the on-board detector proved to be a viable dosimetric quality assurance tool for IMRT with Tomotherapy. Tomotherapy was stable from the dosimetric point of view during the twelve-month period.Entities:
Keywords: Dosimetric stability; Tomotherapy; mega voltage computed tomography detector
Year: 2009 PMID: 20098540 PMCID: PMC2805893 DOI: 10.4103/0971-6203.51933
Source DB: PubMed Journal: J Med Phys ISSN: 0971-6203
Figure 1MVCT Detector
Figure 2A customized water tank for profile measurement with ion chamber.
Figure 3At least three rows of the resolution plugs are clearly visible. In addition, all inserted density plugs are visible from the registration panel. Moreover, the plugs can be individually resolved. The image was found to be free of rings and streak artifacts.
Figure 4The resultant dose profiles (lateral) measured with the ion chamber. The high energy and low-energy beam lateral beam profiles were first measured in water at a depth of 1.5 cm using an A1SL ion chamber. The SSD was 85 cm. The resultant dose profiles (lateral) measured with the on board MVCT detector. The same high-energy and low-energy beams were measured. The couch was positioned such that nothing attenuated the beam except the gantry covers.
Figure 5The water-tank-measured, high-energy beam profile was divided by the water tank measured, low-energy beam profile and normalized to unity at the isocenter. Likewise, the onboard detector measured highenergy profile was divided by the low-energy profile, and normalized to unity at the isocenter. Both plots are shown. Vertical error bars of +/− 0.5% were added to the water tank measured normalized plot.
Figure 6Comparison of CADD measured with PPC05 parallel plate ion chamber, A1SL ion chamber and TLD.