Literature DB >> 20097644

Accuracy and precision of manufacturer-supplied liquid medication administration devices before and after patient education.

Gina Peacock1, Sarah Parnapy, Shelby Raynor, Stewart Wetmore.   

Abstract

OBJECTIVES: To compare the accuracy and precision of a manufacturer-supplied dropper versus a pharmacy-supplied syringe when measured by a trained student pharmacist and to evaluate patients' ability to accurately measure a prescribed amount of medication with these devices before and after counseling on their use.
METHODS: This was a two-phased study with a pretest/posttest design of 125 community pharmacy patrons of Virginia community pharmacies. Deliverable volume was measured by a student pharmacist for nystatin suspension and digoxin elixir using the manufactured-supplied device. Patients at pharmacies were asked to complete a short survey, then asked to measure a specified dose of nystatin suspension and digoxin elixir (using the manufacturer-supplied administration device). The patients repeated this a second time after a student provided education on how to properly use each administration device. The main outcomes measure was whether education on an administration device would provide more measurement accuracy using medications with different physical properties.
RESULTS: For nystatin, 88% and 85.6% of patients measured with an error of greater than or equal to 20% before and after counseling, respectively. For digoxin, the totals were 24.8% before counseling and 4% after. A statistically significant (P < 0.05) difference was observed in the percent measurement error before and after counseling for nystatin and digoxin both for patients who had been educated before the study and for those who had not been counseled.
CONCLUSION: In general, the accuracy of medication dosage devices seems to be most affected by viscosity of the liquid being measured. With education on proper use, the oral syringe provides a statistically and clinically significant improvement in dosing accuracy for viscous medications compared with the manufacturer-supplied dropper. Pharmacists must review with patients the proper way to use a manufacturer- or pharmacy-supplied oral delivery device for appropriate dosing measurements.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20097644     DOI: 10.1331/JAPhA.2010.09006

Source DB:  PubMed          Journal:  J Am Pharm Assoc (2003)        ISSN: 1086-5802


  6 in total

1.  Safeguarding the process of drug administration with an emphasis on electronic support tools.

Authors:  Hanna M Seidling; Anette Lampert; Kristina Lohmann; Julia T Schiele; Alexander J F Send; Diana Witticke; Walter E Haefeli
Journal:  Br J Clin Pharmacol       Date:  2013-09       Impact factor: 4.335

2.  Design of a Precision Medication Dispenser: Preventing Overdose by Increasing Accuracy and Precision of Dosage.

Authors:  Shriya S Srinivasan; Sai N R Kantareddy; Ebenezer A Nkwate; Praveen Meka; Ivy Chang; Nevan C Hanumara; Khalil B Ramadi
Journal:  IEEE J Transl Eng Health Med       Date:  2018-06-19       Impact factor: 3.316

3.  Experiencing antiretroviral adherence: helping healthcare staff better understand adherence to paediatric antiretrovirals.

Authors:  Benjamin R Phelps; Sarah J Hathcock; Jennifer Werdenberg; Gordon E Schutze
Journal:  J Int AIDS Soc       Date:  2010-12-06       Impact factor: 5.396

4.  Ability of Saudi mothers to appropriately and accurately use dosing devices to administer oral liquid medications to their children.

Authors:  Saja Almazrou; Hind Alsahly; Huda Alwattar; Lamya Alturki; Mona Alamri
Journal:  Drug Healthc Patient Saf       Date:  2014-12-22

5.  Children's medicines in Tanzania: a national survey of administration practices and preferences.

Authors:  Lisa V Adams; Sienna R Craig; Elia John Mmbaga; Helga Naburi; Timothy Lahey; Cameron T Nutt; Rodrick Kisenge; Gary J Noel; Stephen P Spielberg
Journal:  PLoS One       Date:  2013-03-06       Impact factor: 3.240

Review 6.  Individual factors increasing complexity of drug treatment-a narrative review.

Authors:  Steffen J Schmidt; Viktoria S Wurmbach; Anette Lampert; Simone Bernard; Walter E Haefeli; Hanna M Seidling; Petra A Thürmann
Journal:  Eur J Clin Pharmacol       Date:  2020-04-01       Impact factor: 2.953

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.