Literature DB >> 20097102

A comparison of stresses in implantation for grafted and plate-and-screw mandible reconstruction.

Tomohisa Nagasao1, Junpei Miyamoto, Tamotsu Tamaki, Hiromasa Kawana.   

Abstract

OBJECTIVE: During oral rehabilitation by dental implantation for mandibles reconstructed with plate and screws, intensified stresses can develop at the implant-bone interfaces and fixation screw-bone interfaces that might cause complications at these interfaces. The present study aims to evaluate the stresses occurring at these sites using the finite element method. STUDY
DESIGN: Ten computer-aided design models simulating 10 mandibles were produced and were termed normal models. Simulation surgery was performed on these normal models, where parts of the models were removed and replaced by fibula or reconstruction plates plus screws. Depending on the replaced part (body [B] or body and symphysis [BS]) and the reconstruction materials (fibula [Fib] or reconstruction plate [Plate]), the modified models were termed B-Fib, B-Plate, BS-Fib, and BS-Plate models, respectively. For each model, an implant was embedded in the first molar region. Stresses occurring at the implant-bone interfaces and fixation screw-bone interfaces on mastication were calculated using the finite element method and compared among the model groups.
RESULTS: The stresses at the implant-bone interfaces showed no statistically significant differences among the 5 model groups. With the B-Plate and BS-Plate models, stresses at the fixation screw-bone interfaces were nearly twice as intense as those at the implant-bone interfaces.
CONCLUSION: If it is allowed that complication risks correlate with stresses, fixation screws are more subject to failure than implants in performing implantation for mandibles reconstructed with a plate and fixation screws. Therefore, the fixation screws deserve special attention in performing oral rehabilitation for such patients. Copyright 2010 Mosby, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20097102     DOI: 10.1016/j.tripleo.2009.10.009

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod        ISSN: 1079-2104


  4 in total

1.  Reconstruction plates used in the surgery for mandibular discontinuity defect.

Authors:  Guk-Jin Seol; Eun-Gyu Jeon; Jong-Sung Lee; So-Young Choi; Jin-Wook Kim; Tae-Geon Kwon; Jun-Young Paeng
Journal:  J Korean Assoc Oral Maxillofac Surg       Date:  2014-12-26

2.  Finite Element Simulation and Additive Manufacturing of Stiffness-Matched NiTi Fixation Hardware for Mandibular Reconstruction Surgery.

Authors:  Ahmadreza Jahadakbar; Narges Shayesteh Moghaddam; Amirhesam Amerinatanzi; David Dean; Haluk E Karaca; Mohammad Elahinia
Journal:  Bioengineering (Basel)       Date:  2016-12-19

3.  Customized mandibular reconstruction plates improve mechanical performance in a mandibular reconstruction model.

Authors:  Ralf Gutwald; Raimund Jaeger; Floor M Lambers
Journal:  Comput Methods Biomech Biomed Engin       Date:  2016-11-25       Impact factor: 1.763

Review 4.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08
  4 in total