An integral quality monitoring system for real-time verification of intensity modulated radiation therapy.

Mohammad K Islam1, Bernhard D Norrlinger, Jason R Smale, Robert K Heaton, Duncan Galbraith, Cary Fan, David A Jaffray.   

Abstract

PURPOSE: To develop an independent and on-line beam monitoring system, which can validate the accuracy of segment-by-segment energy fluence delivery for each treatment field. The system is also intended to be utilized for pretreatment dosimetric quality assurance of intensity modulated radiation therapy (IMRT), on-line image-guided adaptive radiation therapy, and volumetric modulated arc therapy.
METHODS: The system, referred to as the integral quality monitor (IQM), utilizes an area integrating energy fluence monitoring sensor (AIMS) positioned between the final beam shaping device [i.e., multileaf collimator (MLC)] and the patient. The prototype AIMS consists of a novel spatially sensitive large area ionization chamber with a gradient along the direction of the MLC motion. The signal from the AIMS provides a simple output for each beam segment, which is compared in real time to the expected value. The prototype ionization chamber, with a physical area of 22 x 22 cm2, has been constructed out of aluminum with the electrode separations varying linearly from 2 to 20 mm. A calculation method has been developed to predict AIMS signals based on an elementwise integration technique, which takes into account various predetermined factors, including the spatial response function of the chamber, MLC characteristics, beam transmission through the secondary jaws, and field size factors. The influence of the ionization chamber on the beam has been evaluated in terms of transmission, surface dose, beam profiles, and depth dose. The sensitivity of the system was tested by introducing small deviations in leaf positions. A small set of IMRT fields for prostate and head and neck plans was used to evaluate the system. The ionization chamber and the data acquisition software systems were interfaced to two different types of linear accelerators: Elekta Synergy and Varian iX.
RESULTS: For a 10 x 10 cm2 field, the chamber attenuates the beam intensity by 7% and 5% for 6 and 18 MV beams, respectively, without significantly changing the depth dose, surface dose, and dose profile characteristics. An MLC bank calibration error of 1 mm causes the IQM signal of a 3 x 3 cm2 aperture to change by 3%. A positioning error in a single 5 mm wide leaf by 3 mm in 3 X 3 cm2 aperture causes a signal difference of 2%. Initial results for prostate and head and neck IMRT fields show an average agreement between calculation and measurement to within 1%, with a maximum deviation for each of the smallest beam segments to within 5%. When the beam segments of a prostate IMRT field were shifted by 3 mm from their original position, along the direction of the MLC motion, the IQM signals varied, on average, by 2.5%.
CONCLUSIONS: The prototype IQM system can validate the accuracy of beam delivery in real time by comparing precalculated and measured AIMS signals. The system is capable of capturing errors in MLC leaf calibration or malfunctions in the positioning of an individual leaf. The AIMS does not significantly alter the beam quality and therefore could be implemented without requiring recommissioning measurements.

Entities:  

Mesh:

Year:  2009        PMID: 20095254     DOI: 10.1118/1.3250859

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  Process-based quality management for clinical implementation of adaptive radiotherapy.

Authors:  Camille E Noel; Lakshmi Santanam; Parag J Parikh; Sasa Mutic
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

2.  Real-time beam monitoring for error detection in IMRT plans and impact on dose-volume histograms : A multi-center study.

Authors:  Livia Marrazzo; Chiara Arilli; Marlies Pasler; Martijn Kusters; Richard Canters; Luca Fedeli; Silvia Calusi; Marta Casati; Cinzia Talamonti; Gabriele Simontacchi; Lorenzo Livi; Stefania Pallotta
Journal:  Strahlenther Onkol       Date:  2017-12-18       Impact factor: 3.621

3.  High-performance GPU-based rendering for real-time, rigid 2D/3D-image registration and motion prediction in radiation oncology.

Authors:  Jakob Spoerk; Christelle Gendrin; Christoph Weber; Michael Figl; Supriyanto Ardjo Pawiro; Hugo Furtado; Daniella Fabri; Christoph Bloch; Helmar Bergmann; Eduard Gröller; Wolfgang Birkfellner
Journal:  Z Med Phys       Date:  2011-07-22       Impact factor: 4.820

4.  Sensitivity of the IQM and MatriXX detectors in megavolt photon beams.

Authors:  Oluwaseyi Michael Oderinde; Freek du Plessis
Journal:  Rep Pract Oncol Radiother       Date:  2019-08-02

5.  Usability of detecting delivery errors during treatment of prostate VMAT with a gantry-mounted transmission detector.

Authors:  Hirofumi Honda; Masahide Tominaga; Motoharu Sasaki; Masataka Oita; Hiromitsu Kanzaki; Yasushi Hamamoto; Yoshiaki Ishii; Ryuji Yamamoto; Teruhito Mochizuki; Teruhito Kido; Yoshihiro Uto
Journal:  J Appl Clin Med Phys       Date:  2021-05-05       Impact factor: 2.102

6.  Feasibility of a remote, automated daily delivery verification of volumetric-modulated arc therapy treatments using a commercial record and verify system.

Authors:  Jonas D Fontenot
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

7.  Sensitivity Analysis of the Integral Quality Monitoring System® Using Monte Carlo Simulation.

Authors:  Oluwaseyi M Oderinde; F C P du Plessis
Journal:  Comput Math Methods Med       Date:  2017-08-27       Impact factor: 2.238

8.  Beam perturbation characteristics of a 2D transmission silicon diode array, Magic Plate.

Authors:  Ziyad A Alrowaili; Michael L F Lerch; Marco Petasecca; Martin G Carolan; Peter E Metcalfe; Anatoly B Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102

9.  Validation of a method for in vivo 3D dose reconstruction in SBRT using a new transmission detector.

Authors:  Yuji Nakaguchi; Takeshi Ono; Masato Maruyama; Yoshinobu Shimohigashi; Yudai Kai
Journal:  J Appl Clin Med Phys       Date:  2017-06-02       Impact factor: 2.102

10.  Diode-based transmission detector for IMRT delivery monitoring: a validation study.

Authors:  Taoran Li; Q Jackie Wu; Thomas Matzen; Fang-Fang Yin; Jennifer C O'Daniel
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.