| Literature DB >> 2008171 |
U F Rosenow1, M K Islam, H Gaballa, H Rashid.
Abstract
An energy constancy checking method is presented which involves a specially designed wedge-shaped solid phantom in combination with a multiple channel ionization chamber array known as the Thebes device. Once the phantom/beam scanner combination is set up, measurements for all electron energies can be made and evaluated without re-entering the treatment room. This is also valid for the readjustment of beam energies which are found to deviate from required settings. The immediate presentation of the measurements is in the form of crossplots which resemble depth dose profiles. The evaluation of the measured data can be performed using a hand-held calculator, but processing of the measured signals through a PC-type computer is advisable. The method is insensitive to usual fluctuations in beam flatness. The sensitivity and reproducibility of the method are more than adequate. The method may also be used in modified form for photon beams.Entities:
Mesh:
Year: 1991 PMID: 2008171 DOI: 10.1118/1.596719
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071