Energy constancy checking for electron beams using a wedge-shaped solid phantom combined with a beam profile scanner.

U F Rosenow1, M K Islam, H Gaballa, H Rashid.   

Abstract

An energy constancy checking method is presented which involves a specially designed wedge-shaped solid phantom in combination with a multiple channel ionization chamber array known as the Thebes device. Once the phantom/beam scanner combination is set up, measurements for all electron energies can be made and evaluated without re-entering the treatment room. This is also valid for the readjustment of beam energies which are found to deviate from required settings. The immediate presentation of the measurements is in the form of crossplots which resemble depth dose profiles. The evaluation of the measured data can be performed using a hand-held calculator, but processing of the measured signals through a PC-type computer is advisable. The method is insensitive to usual fluctuations in beam flatness. The sensitivity and reproducibility of the method are more than adequate. The method may also be used in modified form for photon beams.

Entities:  

Mesh:

Year:  1991        PMID: 2008171     DOI: 10.1118/1.596719

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Quality assurance of electron and photon beam energy using the BQ-CHECK phantom.

Authors:  Richard J Speight; Ashraf Esmail; Steve J Weston
Journal:  J Appl Clin Med Phys       Date:  2011-02-01       Impact factor: 2.102

2.  Electron energy constancy verification using a double-wedge phantom.

Authors:  Derek M Wells; Philip J Picco; Will Ansbacher
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

3.  Monitoring beam-quality constancy considering uncertainties associated with ionization chambers in Daily QA3 device.

Authors:  Su Chul Han; Jihun Kim; Min Cheol Han; Kyung Hwan Chang; Kwangwoo Park; Ho Jin Kim; Dong Wook Kim; Jin Sung Kim
Journal:  PLoS One       Date:  2021-02-17       Impact factor: 3.240

4.  Detection of electron beam energy variations using a computed radiography system.

Authors:  Yang C Cai; Yuanyuan Ge; Damian Bernard; Julius Turian; James C H Chu
Journal:  J Appl Clin Med Phys       Date:  2009-10-15       Impact factor: 2.102

5.  Electron beam energy QA - a note on measurement tolerances.

Authors:  Juergen Meyer; Matthew J Nyflot; Wade P Smith; Landon S Wottoon; Lori Young; Fei Yang; Minsun Kim; Kristi R G Hendrickson; Eric Ford; Alan M Kalet; Ning Cao; Claire Dempsey; George A Sandison
Journal:  J Appl Clin Med Phys       Date:  2016-03-08       Impact factor: 2.102