Evaluation of the sensitivity of an in vitro high frequency ultrasound device to monitor the coagulation process: study of the effects of heparin treatment in a murine model.

Rachel Callé1, Gaël Y Rochefort, Nicolas Desbuards, Camille Plag, Daniel Antier, Frédéric Ossant.   

Abstract

This study evaluates the sensitivity of a new in vitro high frequency ultrasound test of the whole blood coagulation process. A rat model of anticoagulant treatment is reported. Many recent studies of the role of red blood cells in the whole blood coagulation process have revealed an increasing demand for global tests of the coagulation process performed on whole blood instead of plasma samples. In contrast to existing optical tests, high frequency ultrasound presents the advantages of characterizing the mechanical properties of whole blood clotting. Ultrasound longitudinal wave velocity and integrated attenuation coefficient (IAC) were simultaneously assessed in a 10 to 30 MHz frequency range during the whole blood coagulation process in vitro in rats under anticoagulant therapy. Differences between humans and rats were also clearly emphasized in non-clotting blood and in clotting blood using specific criteria deduced from acoustic parameters (ultrasound velocity for non-clotting blood:=1574+/-2m/s for rats and 1583+/-3m/s for humans and IAC=2.25+/-0.14 dB/cm for rats and 1.5+/-0.23 dB/cm for humans). We also measured the coagulation time t(0) from the acoustic velocity (t(0) =11.15+/-7 min for control rat blood and 43.3+/-11.4 min for human blood). Different doses of heparin were administered to rats. The sensitivity of the ultrasound device to the effects of heparin was evaluated. Differences between non-treated rats and chronically and acutely treated rats were recorded and quantified. We particularly noted that the slope S and the amplitude I of the variations in acoustic velocity were linked to clot retraction, which is a good indicator of the platelet function. The amplitude of the variations in S was between (20+/-8) x1 0(-3) m/s(2) for control group rats, and (0.92+/-0.35) x 10(-3) m/s(2) for chronic heparin-treated group rats. The values of I were 15 times higher for control group rats than for chronic heparin-treated group rats. 2010 World Federation for Ultrasound in Medicine & Biology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20045589     DOI: 10.1016/j.ultrasmedbio.2009.10.010

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  1 in total

1.  A novel ultrasonic method for evaluation of blood clotting parameters.

Authors:  Vacis Tatarunas; Algirdas Voleisis; Reimondas Sliteris; Rymantas Kazys; Liudas Mazeika; Vaiva Lesauskaite
Journal:  J Med Ultrason (2001)       Date:  2018-03-13       Impact factor: 1.314

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.