Can the Bio-Transfx pin fail during initial ACL graft insertion?

Alexander Golant1, Eric J Strauss, Kevin Khajavi, Orrin H Sherman, Jeffrey E Rosen.   

Abstract

BACKGROUND: The Bio-Transfix pin is a biodegradable device used for femoral tunnel anterior cruciate ligament (ACL) graft fixation. Recent clinical studies have suggested the possibility of the pin's postoperative failure.
METHODS: This investigation evaluates the initial strength of several Bio-Transfix pin ACL fixations in a simulated femoral tunnel model. The forces generated by five surgeons during simulated ACL graft tensioning were also measured.
RESULTS: Average strengths of the pins ranged from 1075 to 2160 N for 10 and 8 mm tunnels, respectively, whereas the maximum surgeon-generated forces were 535 N.
CONCLUSIONS: These results imply that initial fracture of the pin itself is unlikely; however, failure of the supporting bone or a decrease in pin strength due to biodegradation could account for early loss of the fixation.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 20001934

Source DB:  PubMed          Journal:  Bull NYU Hosp Jt Dis        ISSN: 1936-9719


  2 in total

1.  Anterior cruciate ligament reconstruction using biodegradable transfemoral fixation at 5-year follow-up: clinical and magnetic resonance imaging evaluation.

Authors:  Sven Nebelung; Gregor Deitmer; Rolf Gebing; Frank Reichwein; Wolfgang Nebelung
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2012-03-06       Impact factor: 4.342

2.  Breakage of bioabsorbable femoral transfix screw after arthroscopic anterior cruciate ligament repair: a report of two cases.

Authors:  A Raviraj; Ashish Anand; K Santosh
Journal:  Eur J Orthop Surg Traumatol       Date:  2012-02-04
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.