Literature DB >> 19969365

Comparison of the diagnostic capability of the Heidelberg Retina Tomographs 2 and 3 for glaucoma in the Indian population.

Harsha L Rao1, Ganesh J Babu, Garudadri C Sekhar.   

Abstract

PURPOSE: To compare the diagnostic capability of the Heidelberg Retina Tomograph (HRT) 2 and 3 (Heidelberg Engineering, GmBH, Dossenheim, Germany) for glaucoma in an Indian population.
DESIGN: Evaluation of a diagnostic technology. PARTICIPANTS: Ninety-eight glaucoma subjects and 79 normal controls.
METHODS: All participants underwent imaging with HRT2. Heidelberg Retina Tomograph 2 examinations were exported to HRT3 software. The stereometric parameters of HRT2 and HRT3 were compared. The diagnostic capability of Moorfields Regression Analysis (MRA) in the HRT2, HRT3 with and without ethnicity correction, and Glaucoma Probability Score (GPS) of HRT3 were compared. Analysis was done with the borderline results of MRA and GPS considered as normal to derive the most specific criteria and as abnormal to derive the least specific criteria. MAIN OUTCOME MEASURES: Sensitivity, specificity, and area under the receiver operating characteristic curve (AUC).
RESULTS: On standard automated perimetry (SAP), the mean deviation (mean+/-standard deviation) of the glaucoma and normal groups were -7.3+/-6.7 dB and -0.4+/-1.1 dB, respectively (P<0.001). The AUC for the individual stereometric parameters of HRT2 were similar to that of HRT3. The sensitivity and specificity (most specific criteria) were 56.1% and 96.2%, respectively, for MRA of HRT2 and 56.1% and 91.1%, respectively, for MRA of HRT3. The sensitivity and specificity (least specific criteria) were 88.8% and 88.6%, respectively, for MRA of HRT2 and 70.4% and 81%, respectively, for MRA of HRT3. By using the Indian specific normative database of HRT3, the sensitivity and specificity were 37.8% and 97.5%, respectively, for the most specific criteria, and 55.1% and 89.9%, respectively, for the least specific criteria. Sensitivity and specificity of GPS were 86.7% and 68.3%, respectively, for the most specific criteria, and 93.9% and 39.2%, respectively, for the least specific criteria.
CONCLUSIONS: In this cohort, the diagnostic capability of HRT2 MRA was similar to that of HRT3 MRA; ethnicity correction did not improve results; GPS was more sensitive but less specific than MRA in diagnosing glaucoma. Copyright (c) 2010 American Academy of Ophthalmology. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19969365     DOI: 10.1016/j.ophtha.2009.06.062

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  14 in total

1.  Influence of clinically invisible, but optical coherence tomography detected, optic disc margin anatomy on neuroretinal rim evaluation.

Authors:  Alexandre S C Reis; Neil O'Leary; Hongli Yang; Glen P Sharpe; Marcelo T Nicolela; Claude F Burgoyne; Balwantray C Chauhan
Journal:  Invest Ophthalmol Vis Sci       Date:  2012-04-18       Impact factor: 4.799

2.  Normative Databases for Imaging Instrumentation.

Authors:  Tony Realini; Linda M Zangwill; John G Flanagan; David Garway-Heath; Vincent M Patella; Chris A Johnson; Paul H Artes; Ian B Gaddie; Murray Fingeret
Journal:  J Glaucoma       Date:  2015-08       Impact factor: 2.503

3.  Effect of race, age, and axial length on optic nerve head parameters and retinal nerve fiber layer thickness measured by Cirrus HD-OCT.

Authors:  O'Rese J Knight; Christopher A Girkin; Donald L Budenz; Mary K Durbin; William J Feuer
Journal:  Arch Ophthalmol       Date:  2012-03

4.  Bruch's Membrane Opening Minimum Rim Width and Retinal Nerve Fiber Layer Thickness in a Normal White Population: A Multicenter Study.

Authors:  Balwantray C Chauhan; Vishva M Danthurebandara; Glen P Sharpe; Shaban Demirel; Christopher A Girkin; Christian Y Mardin; Alexander F Scheuerle; Claude F Burgoyne
Journal:  Ophthalmology       Date:  2015-07-18       Impact factor: 12.079

5.  Structural and functional changes to the retina and optic nerve following panretinal photocoagulation over a 2-year time period.

Authors:  R Filek; P Hooper; T Sheidow; J Gonder; D K Varma; L Heckler; W Hodge; S Chakrabarti; C M L Hutnik
Journal:  Eye (Lond)       Date:  2017-04-28       Impact factor: 3.775

Review 6.  Optic nerve head and fibre layer imaging for diagnosing glaucoma.

Authors:  Manuele Michelessi; Ersilia Lucenteforte; Francesco Oddone; Miriam Brazzelli; Mariacristina Parravano; Sara Franchi; Sueko M Ng; Gianni Virgili
Journal:  Cochrane Database Syst Rev       Date:  2015-11-30

7.  Diagnosing preperimetric glaucoma with spectral domain optical coherence tomography.

Authors:  Renato Lisboa; Mauro T Leite; Linda M Zangwill; Ali Tafreshi; Robert N Weinreb; Felipe A Medeiros
Journal:  Ophthalmology       Date:  2012-08-09       Impact factor: 12.079

8.  Diagnostic Accuracy of the Spectralis and Cirrus Reference Databases in Differentiating between Healthy and Early Glaucoma Eyes.

Authors:  Anna L Silverman; Naama Hammel; Naira Khachatryan; Lucie Sharpsten; Felipe A Medeiros; Christopher A Girkin; Jeffrey M Liebmann; Robert N Weinreb; Linda M Zangwill
Journal:  Ophthalmology       Date:  2015-10-31       Impact factor: 12.079

9.  Protruded retinal layers within the optic nerve head neuroretinal rim.

Authors:  Lucas A Torres; Jayme R Vianna; Faisal Jarrar; Glen P Sharpe; Makoto Araie; Joseph Caprioli; Shaban Demirel; Christopher A Girkin; Masanori Hangai; Aiko Iwase; Jeffrey M Liebmann; Christian Y Mardin; Toru Nakazawa; Harry A Quigley; Alexander F Scheuerle; Kazuhisa Sugiyama; Hidenobu Tanihara; Goji Tomita; Yasuo Yanagi; Claude F Burgoyne; Balwantray C Chauhan
Journal:  Acta Ophthalmol       Date:  2018-06       Impact factor: 3.761

10.  Evidence-based approach to glaucoma management.

Authors:  Chandrasekhar Garudadri; Sirisha Senthil; Harsha Laxmana Rao
Journal:  Indian J Ophthalmol       Date:  2011-01       Impact factor: 1.848

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.