A 55-year-old male patient presented with an acute myocardial infarction. A sirolimus-eluting stent (SES) was implanted in the proximal left anterior descending artery (LAD). Eight months later, there was a newly developed distal LAD lesion. An additional SES was implanted. Twenty-eight months after the index procedure of primary coronary intervention, the electrocardiogram showed ST elevation in the precordial leads and an emergency coronary angiogram showed diffuse stent thrombosis (ST) in the proximal LAD. Thirty-four months after the index procedure, coronary angiography showed a large peri-stent coronary aneurysm in the proximal LAD and focal in-stent restenosis (ISR) at the proximal edge of the distal LAD stent. On fluoroscopy, a fracture was noted in the middle part of the distal SES. A zotarolimus- eluting stent (ZES) was deployed and overlapped the restenosis and fracture sites. Forty months after the index procedure, there were no changes in the size of the aneurysm or in the other stent complications including the fracture and restenosis. At present, the patient has remained asymptomatic for eight months.Entities:
Keywords: Drug-eluting stents; Percutaneous transluminal coronary augioplasty
Year: 2009 PMID: 19949591 PMCID: PMC2771792 DOI: 10.4070/kcj.2009.39.10.439
Source DB: PubMed Journal: Korean Circ J ISSN: 1738-5520 Impact factor: 3.243
Fig. 1Coronary angiography shows a stent thrombosis in the proximal left anterior descending artery (arrow) (A), post-balloon angioplasty (B). An intravascular ultrasound shows late stent malapposition (*) (C). Angiography after five days reveals restoration of the normal blood flow (D).
Fig. 2Follow-up angiography reveals a large proximal left anterior descending artery aneurysm (arrow) (A) and an intravascular ultrasound shows malapposition of the stent (B).
Fig. 3Coronary angiography demonstrates in-stent restenosis (white arrow) (A), while fluoroscopy shows a fractured sirolimus-eluting stent (black arrow) (B).