The proximal fixation strength of modern EVAR grafts in a short aneurysm neck. An in vitro study.

W M P F Bosman1, T J V D Steenhoven, D R Suárez, J W Hinnen, E R Valstar, J F Hamming.   

Abstract

OBJECTIVES: The study aims to measure the strength of the proximal fixation of endografts in short and long necks.
DESIGN: Three types of endografts were compared: Gore Excluder, Vascutek Anaconda and Medtronic Endurant.
MATERIALS AND METHODS: The proximal part of the stent grafts was inserted in bovine arteries and the graft was then attached to a tensile testing machine. The force to obtain dislodgement (DF) from the aorta was recorded for each graft at proximal seal lengths of 10 and 15 mm.
RESULTS: The median DF (interquartile range, IQR) for the Excluder, the Anaconda and the Endurant with a seal length of 15 mm was: 11.8 (10.5-12.0) N, 20.8 (18.0-30.1) N and 10.7 (10.4-11.3) N. With the shorter proximal seal of 10mm, DF was, respectively: 6.0 (4.5-6.6) N, 17.0 (11.2-36.6) N and 6.4 (6.1-12.0) N.
CONCLUSIONS: The proximal fixation of the Anaconda is superior to the Excluder and the Endurant at short necks of 10 and 15 mm in an experimental set-up. There is a statistically significant decrease of proximal fixation for the Excluder stent graft, when decreasing the length of the proximal neck from 15 to 10 mm. Copyright (c) 2009 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19939708     DOI: 10.1016/j.ejvs.2009.10.019

Source DB:  PubMed          Journal:  Eur J Vasc Endovasc Surg        ISSN: 1078-5884            Impact factor:   7.069


  6 in total

1.  Surgical stent planning: simulation parameter study for models based on DICOM standards.

Authors:  S Scherer; T Treichel; N Ritter; G Triebel; W G Drossel; O Burgert
Journal:  Int J Comput Assist Radiol Surg       Date:  2010-07-22       Impact factor: 2.924

2.  Stent graft visualization and planning tool for endovascular surgery using finite element analysis.

Authors:  S von Sachsen; B Senf; O Burgert; J Meixensberger; H J Florek; F W Mohr; C D Etz
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-10-30       Impact factor: 2.924

Review 3.  A critical appraisal of endovascular stent-grafts in the management of abdominal aortic aneurysms.

Authors:  Nikolaos Schoretsanitis; Efstratios Georgakarakos; Christos Argyriou; Kiriakos Ktenidis; George S Georgiadis
Journal:  Radiol Med       Date:  2017-01-21       Impact factor: 3.469

4.  Stimulation of the dorsal root ganglion using an Injectrode®.

Authors:  Ashley N Dalrymple; Jordyn E Ting; Rohit Bose; James K Trevathan; Stephan Nieuwoudt; Scott F Lempka; Manfred Franke; Kip A Ludwig; Andrew J Shoffstall; Lee E Fisher; Douglas J Weber
Journal:  J Neural Eng       Date:  2021-11-04       Impact factor: 5.379

5.  An in Vitro Twist Fatigue Test of Fabric Stent-Grafts Supported by Z-Stents vs. Ringed Stents.

Authors:  Jing Lin; Robert Guidoin; Jia Du; Lu Wang; Graeham Douglas; Danjie Zhu; Mark Nutley; Lygia Perron; Ze Zhang; Yvan Douville
Journal:  Materials (Basel)       Date:  2016-02-16       Impact factor: 3.623

6.  Comparative evaluation of ballet-type and conventional stent graft configurations for endovascular aneurysm repair: A CFD analysis.

Authors:  Fahmida Ashraf; Tehmina Ambreen; Cheol Woo Park; Dong-Ik Kim
Journal:  Clin Hemorheol Microcirc       Date:  2021       Impact factor: 2.375

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.