Literature DB >> 19926777

Six of eight hemoglobin A1c point-of-care instruments do not meet the general accepted analytical performance criteria.

Erna Lenters-Westra1, Robbert J Slingerland.   

Abstract

BACKGROUND: Hemoglobin A(1c) (Hb A(1c)) point-of-care (POC) instruments are widely used to provide rapid-turnaround results in diabetic care centers. We investigated the conformance of various Hb A(1c) POC instruments (In2it from Bio-Rad, DCA Vantage from Siemens, Afinion and Nycocard from Axis-Shield, Clover from Infopia, InnovaStar from DiaSys, A1CNow from Bayer, and Quo-Test from Quotient Diagnostics) with generally accepted performance criteria for Hb A(1c).
METHODS: The CLSI protocols EP-10, EP-5, and EP-9 were applied to investigate imprecision, accuracy, and bias. We assessed bias using 3 certified secondary reference measurement procedures and the mean of the 3 reference methods. Assay conformance with the National Glycohemoglobin Standardization Program (NGSP) certification criteria, as calculated from analyses with 2 different reagent lot numbers for each Hb A(1c) method, was also evaluated.
RESULTS: Because of disappointing EP-10 results, 2 of the 8 manufacturers decided not to continue the evaluation. The total CVs from EP-5 evaluations for the different instruments with a low and high Hb A(1c) value were: In2it 4.9% and 3.3%, DCA Vantage 1.8% and 3.7%, Clover 4.0% and 3.5%, InnovaStar 3.2% and 3.9%, Nycocard 4.8% and 5.2%, and Afinion 2.4% and 1.8%. Only the Afinion and the DCA Vantage passed the NGSP criteria with 2 different reagent lot numbers.
CONCLUSIONS: Only the Afinion and the DCA Vantage met the acceptance criteria of having a total CV <3% in the clinically relevant range. The EP-9 results and the calculations of the NGSP certification showed significant differences in analytical performance between different reagent lot numbers for all Hb A(1c) POC instruments.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19926777     DOI: 10.1373/clinchem.2009.130641

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  60 in total

Review 1.  Molecular analysis of blood with micro-/nanoscale field-effect-transistor biosensors.

Authors:  Matthew S Makowski; Albena Ivanisevic
Journal:  Small       Date:  2011-06-03       Impact factor: 13.281

2.  Analysis of the accuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.

Authors:  Randie R Little
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

Review 3.  Point-of-Care Hemoglobin A1c Testing: An Evidence-Based Analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2014-07-01

4.  Accuracy and precision of the Axis-Shield Afinion hemoglobin A1c measurement device.

Authors:  Jamie R Wood; Brett M Kaminski; Craig Kollman; Roy W Beck; Callyn A Hall; Jason P Yun; Eda Cengiz; Michael J Haller; Krishna Hassan; Georgeanna J Klingensmith; William V Tamborlane
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

5.  Analytical performance and clinical use of a hemoglobin A1c point-of-care analyzer in a pediatric unit.

Authors:  Anne Marie Dupuy; Cécile Elong-Bertard; Stéphanie Badiou; Amandine Barrot; Anne Sophie Bargnoux; Jean Paul Cristol
Journal:  J Diabetes Sci Technol       Date:  2013-09-01

6.  Comparison of Arkray/ELITech ADAMS HA-8180V with Bio-Rad Variant, II Turbo2.0 and Tosoh Bioscience HLC-723G8 for HbA1c determination.

Authors:  Stéphanie Badiou; Julie Guillot; Nils Kuster; Anne-Sophie Bargnoux; Patricia Aguilar-Martinez; Elodie Boissier; Jean-Paul Cristol; Anne-Marie Dupuy
Journal:  J Clin Lab Anal       Date:  2014-03-20       Impact factor: 2.352

7.  Point-of-care testing in hospitals and primary care: Unsatisfactory quality.

Authors:  Matthias Orth
Journal:  Dtsch Arztebl Int       Date:  2011-02-25       Impact factor: 5.594

8.  Primary care diagnostic technology update: point-of-care testing for glycosylated haemoglobin.

Authors:  Annette Plüddemann; Christopher P Price; Matthew Thompson; Jane Wolstenholme; Carl Heneghan
Journal:  Br J Gen Pract       Date:  2011-02       Impact factor: 5.386

9.  Retrospective analysis of 55,769 HbA1c EQA results obtained from professional laboratories and medical offices participating in surveys organized by two European EQA centers over a nine-year period.

Authors:  Pierre-Alain Morandi; André Deom; Dagmar Kesseler; Richard Cohen
Journal:  J Clin Lab Anal       Date:  2011       Impact factor: 2.352

10.  Performance of hemoglobin A1c assay methods: good enough?

Authors:  Randie R Little
Journal:  Clin Chem       Date:  2014-06-17       Impact factor: 8.327

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.