Literature DB >> 19924775

Transcatheter pulmonary valve implantation using the Edwards SAPIEN transcatheter heart valve.

Robert H Boone1, John G Webb, Eric Horlick, Lee Benson, Qi-Ling Cao, N Nadeem, Marla Kiess, Ziyad M Hijazi.   

Abstract

BACKGROUND: Conduits placed in the right ventricular outflow tract (RVOT) have limited longevity which often requires increasingly complex reoperations. Transcatheter pulmonary valve implantation improves conduit hemodynamics through a minimally invasive approach. We present data for 7 patients treated with the Edwards SAPIEN transcatheter heart valve (THV). PATIENTS: Patients' ranged in age from 16 to 52 years, one was female, and all had NYHA class II-III symptoms. Patients had pulmonary homografts that had been placed 2-25 years earlier during the Ross procedure (n = 4), repaired double outlet right ventricle with situs inversus (n = 1), or Rastelli repair for D-TGA, pulmonary atresia, and ventricular septal defect (n = 2). Patients had either severe pulmonary stenosis and/or moderate to severe pulmonary regurgitation.
RESULTS: All patients had successful percutaneous implantation of the 23 mm SAPIEN THV under general anesthesia. Fluoroscopy times ranged from 16 to 49 mins and procedure times ranged from 110 to 237 mins. The RV:systemic pressure ratio decreased from 78 +/- 18 to 39 +/- 8%, the RVOT gradient improved from 60.7 +/- 24.3 to 14.9 +/- 6.9 mm Hg, no patients had pulmonary insufficiency, and all patients had symptom improvement. At a maximum follow-up of 3.5 years (median 22.5 months), Doppler peak gradients ranged from 7-36 mm Hg, and there is no evidence of late stent fracture or structural valve failure.
CONCLUSION: The SAPIEN THV can be used successfully in the treatment of patients with right ventricle to pulmonary artery homograft failure. The valve is durable to at least 3.5 years without stent fracture or regurgitation. Clinical trials are underway to assess the long-term safety and efficacy of this valve.

Entities:  

Mesh:

Year:  2010        PMID: 19924775     DOI: 10.1002/ccd.22250

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  14 in total

Review 1.  [Transcatheter valve interventions for heart valve diseases].

Authors:  A Schaefer; H Bertram
Journal:  Internist (Berl)       Date:  2010-12       Impact factor: 0.743

2.  Evaluation of Hybrid Surgical Access Approaches for Pulmonary Valve Implantation in an Acute Swine Model.

Authors:  Ruth Thalmann; Elena M Merkel; Bassil Akra; Rene Bombien; Rainer G Kozlik-Feldmann; Christoph Schmitz
Journal:  Comp Med       Date:  2019-06-20       Impact factor: 0.982

Review 3.  Valve-in-valve implantations: is this the new standard for degenerated bioprostheses? Review of the literature.

Authors:  Krys Milburn; Vinayak Bapat; Martyn Thomas
Journal:  Clin Res Cardiol       Date:  2014-01-21       Impact factor: 5.460

4.  Percutaneous implantation of the Edwards SAPIEN(™) pulmonic valve: initial results in the first 22 patients.

Authors:  Nikolaus A Haas; Axel Moysich; Ulrich Neudorf; Hojjat Mortezaeian; Mohamed Abdel-Wahab; Heike Schneider; Daniel De Wolf; Jerome Petit; Sreeram Narayanswami; Kai Thorsten Laser; Eugen Sandica
Journal:  Clin Res Cardiol       Date:  2012-08-30       Impact factor: 5.460

Review 5.  Will catheter interventions replace surgery for valve abnormalities?

Authors:  Michael L O'Byrne; Matthew J Gillespie
Journal:  Curr Opin Cardiol       Date:  2014-01       Impact factor: 2.161

6.  Percutaneous pulmonary and tricuspid valve implantations: An update.

Authors:  Robert Wagner; Ingo Daehnert; Philipp Lurz
Journal:  World J Cardiol       Date:  2015-04-26

7.  Transcatheter Pulmonary Valve Replacement in a Carcinoid Heart.

Authors:  Pranav Loyalka; Michael Schechter; Angelo Nascimbene; Ajay Sundara Raman; Cezar A Ilieascu; Igor D Gregoric; Biswajit Kar
Journal:  Tex Heart Inst J       Date:  2016-08-01

8.  [Minimally invasive pulmonary valve replacement in pediatric patients: importance of imaging].

Authors:  K Bauner; R Kozlik-Feldmann
Journal:  Radiologe       Date:  2013-10       Impact factor: 0.635

9.  Percutaneous pulmonary valve implantation.

Authors:  Young-Seok Lee; Hyoung-Doo Lee
Journal:  Korean Circ J       Date:  2012-10-31       Impact factor: 3.243

Review 10.  Interventional cardiology in adults with congenital heart disease.

Authors:  Harsimran S Singh; Eric Horlick; Mark Osten; Lee N Benson
Journal:  Nat Rev Cardiol       Date:  2013-08-27       Impact factor: 32.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.