LCD versus CRT monitors for digital mammography: a comparison of observer performance for the detection of clustered microcalcifications and masses.

Joo Hee Cha1, Woo Kyung Moon, Nariya Cho, Eun Hye Lee, Jeong Seon Park, Mi Jung Jang.   

Abstract

BACKGROUND: There are two types of soft-copy display of digital mammograms, cathode-ray tube (CRT) and liquid-crystal display (LCD) monitors, and these present different resolutions and noise characteristics.
PURPOSE: To compare observer performances for the detection of clustered microcalcifications and masses in digital mammograms, using LCD and CRT monitors.
MATERIAL AND METHODS: Four radiologists evaluated 100 digital mammograms. Of these, 28 mammograms depicted clustered microcalcifications (12 benign, 16 malignant), 18 depicted masses (10 benign, eight malignant), and 54 depicted no apparent abnormality. The images were stored uncompressed as Digital Imaging and Communications in Medicine (DICOM) files and randomized for two sessions of soft-copy reading. The readers independently read all cases displayed on two 5-megapixel LCD and two CRT monitors with an interval of 3 months. Observers were asked to rate using a five-point scale the likelihoods of microcalcifications and masses independently. Receiver operating characteristic analysis was used to analyze results.
RESULTS: For the LCD and CRT monitors, average area-under-the-curve (A(z)) values for microcalcification detection were 0.977 (0.976, 0.970, 0.980, and 0.983 for the four readers) and 0.958 (0.910, 0.990, 0.976, 0.956), respectively. Corresponding A(z) values for masses were 0.971 (0.955, 0.990, 0.949, 0.989) and 0.944 (0.928, 0.966, 0.955, 0.925). No statistically significant difference was detected between the two monitor types (P>0.05).
CONCLUSION: The LCD and CRT monitors are comparable in terms of detecting clustered microcalcifications and masses in digital mammograms.

Entities:  

Mesh:

Year:  2009        PMID: 19922305     DOI: 10.3109/02841850903246608

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  2 in total

1.  Effect of greyscale liquid crystal displays of different resolutions on observer performance during detection of small solitary pulmonary nodules.

Authors:  J Yin; Q Guo; W Zhang; H Su; J Zhang; Y Yue; C Ding; A Lin; Y Wang; H Wang
Journal:  Br J Radiol       Date:  2012-06-27       Impact factor: 3.039