Accuracy of WASCA aberrometer refraction compared to manifest refraction in Chinese adult myopes.

Xiangjia Zhu1, Jinhui Dai, Renyuan Chu, Yi Lu, Xingtao Zhou, Lin Wang.   

Abstract

PURPOSE: To investigate the accuracy of the Carl Zeiss Meditec Wavefront Supported Custom Ablation (WASCA) aberrometer in refraction testing for Chinese myopic adults.
METHODS: Manifest refraction and WASCA were performed on 360 eyes of 360 consecutive Chinese myopic adults without cycloplegia. Both vector analysis and conventional notation were applied for comparing results, and differences between WASCA and manifest refraction for each component were calculated for accuracy evaluation. Correlation coefficients, regression equations, and mean errors of each component between the two methods were also tested.
RESULTS: Pearson correlation coefficient for M, J(0), J(45), sphere, and cylinder was 0.9680, 0.9320, 0.8655, 0.9668, and 0.8761, respectively. Mean error for the components above was -0.39+/-0.71 diopters (D), -0.02+/-0.16 D, 0.01+/-0.12 D, -0.42+/-0.70 D, and 0.06+/-0.30 D, respectively. In patients whose manifest vector M was < or =-6.00 D, mean errors of sphere and vector M were larger than the remainder of the study group. Ninety percent cylinder errors were within +/-0.50 D. Mean error of axis was 7.4 degrees, with 50% of eyes within 5 degrees. In cases with < or =-0.75 D astigmatism, mean error of axis was 9.8 degrees, which was larger than the remainder of the study group by 4.9 degrees.
CONCLUSIONS: For Chinese myopic adults, the concordance between WASCA and manifest refraction is, on average, high. However, for eyes with low to moderate spherocylindrical refraction, WASCA is less accurate in predicting sphere and vector M. It is also less effective for the measurement of axis in patients with astigmatism <0.75 D. Copyright 2009, SLACK Incorporated.

Entities:  

Mesh:

Year:  2009        PMID: 19921772     DOI: 10.3928/1081597X-20091016-09

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  8 in total

1.  Accuracy of wavefront aberrometer refraction vs manifest refraction in cataract patients: impact of age, ametropia and visual function.

Authors:  Jan O Huelle; Toam Katz; Jan Draeger; Milena Pahlitzsch; Vasyl Druchkiv; Johannes Steinberg; Gisbert Richard; Stephan J Linke
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-01-05       Impact factor: 3.117

2.  Is an objective refraction optimised using the visual Strehl ratio better than a subjective refraction?

Authors:  Gareth D Hastings; Jason D Marsack; Lan Chi Nguyen; Han Cheng; Raymond A Applegate
Journal:  Ophthalmic Physiol Opt       Date:  2017-03-30       Impact factor: 3.117

3.  Effect of pupil size on higher-order aberrations in high-myopic pseudophakic eyes with posterior staphyloma.

Authors:  X Zhu; H Ye; J Yang; Y Lu
Journal:  Eye (Lond)       Date:  2014-10-17       Impact factor: 3.775

4.  Clinical efficacy of implantation of toric intraocular lenses with different incision positions: a comparative study of steep-axis incision and non-steep-axis incision.

Authors:  Wenwen He; Xiangjia Zhu; Yu Du; Jin Yang; Yi Lu
Journal:  BMC Ophthalmol       Date:  2017-07-27       Impact factor: 2.209

5.  Repeatability of Aberrometry-Based Automated Subjective Refraction in Healthy and Keratoconus Subjects.

Authors:  Gonzalo Carracedo; Carlos Carpena-Torres; Cristina Pastrana; Ana Privado-Aroco; María Serramito; Laura Batres
Journal:  J Ophthalmol       Date:  2020-10-30       Impact factor: 1.909

6.  Agreement of wavefront-based refraction, dry and cycloplegic autorefraction with subjective refraction.

Authors:  Shahram Bamdad; Hamed Momeni-Moghaddam; Milad Abdolahian; David P Piñero
Journal:  J Optom       Date:  2020-09-28

7.  Comparison of Two Wavefront Autorefractors: Binocular Open-Field versus Monocular Closed-Field.

Authors:  Gonzalo Carracedo; Carlos Carpena-Torres; Laura Batres; Maria Serramito; Anahí Gonzalez-Bergaz
Journal:  J Ophthalmol       Date:  2020-01-03       Impact factor: 1.909

8.  Accuracy of WASCA Aberrometer Refraction Compared to Manifest Refraction and Cycloplegic Refraction in Hyperopia Measurement.

Authors:  Dan Fu; Xuan Ding; Jianmin Shang; Zhiqiang Yu; Xingtao Zhou
Journal:  Transl Vis Sci Technol       Date:  2020-10-06       Impact factor: 3.283

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.