Literature DB >> 33489331

Repeatability of Aberrometry-Based Automated Subjective Refraction in Healthy and Keratoconus Subjects.

Gonzalo Carracedo1, Carlos Carpena-Torres1, Cristina Pastrana1, Ana Privado-Aroco1, María Serramito1, Laura Batres1.   

Abstract

PURPOSE: To compare the intersession repeatability of the Eye Refract, a new instrument to perform aberrometry-based automated subjective refraction, on healthy and keratoconus subjects.
MATERIALS AND METHODS: A cross-sectional and randomized study was performed. A total of 64 participants were evaluated in the study, selecting one eye per participant randomly. The sample was divided into two different groups: 33 healthy subjects (38.85 ± 13.21 years) and 31 with keratoconus (37.29 ± 11.37 years). Three refractions per participant with the Eye Refract were performed on three different days, without cycloplegia. The repeatability analysis of refractive variables (M, J0, and J45), binocular corrected distance visual acuity (BCDVA), and spent time in refraction was performed in terms of repeatability (S r), its 95% confidence interval (r), and intraclass correlation coefficient (ICC).
RESULTS: There were no statistically significant differences (P ≥ 0.05) between sessions in both groups for all refractive variables (M, J0, and J45) and BCDVA. Spent time in refraction was reduced as the sessions went by (P < 0.05). The Eye Refract was more repeatable for refractive errors assessment in healthy subjects (M : S r  = 0.27 D; J0 : S r  = 0.09 D; J45 : S r  = 0.06 D) compared to those with keratoconus (M : S r  = 0.65 D; J0 : S r  = 0.29 D; J45 : S r  = 0.24 D), while it was similar for BCDVA.
CONCLUSIONS: The Eye Refract offered better repeatability to assess refractive errors in healthy subjects compared to those with keratoconus. Despite measurements being also consistent in keratoconus subjects, they should be treated with caution in clinical practice.
Copyright © 2020 Gonzalo Carracedo et al.

Entities:  

Year:  2020        PMID: 33489331      PMCID: PMC7803284          DOI: 10.1155/2020/4831298

Source DB:  PubMed          Journal:  J Ophthalmol        ISSN: 2090-004X            Impact factor:   1.909


  30 in total

1.  Clinical evaluation of the L80 autorefractometer.

Authors:  Einat Shneor; Michel Millodot; Ortal Avraham; Shany Amar; Ariela Gordon-Shaag
Journal:  Clin Exp Optom       Date:  2011-09-20       Impact factor: 2.742

2.  Precision (repeatability and reproducibility) studies and sample-size calculation.

Authors:  Colm McAlinden; Jyoti Khadka; Konrad Pesudovs
Journal:  J Cataract Refract Surg       Date:  2015-12       Impact factor: 3.351

3.  Visual acuity repeatability in keratoconus: impact on sample size. Collaborative Longitudinal Evaluation of Keratoconus (CLEK) Study Group.

Authors:  M O Gordon; K B Schechtman; L J Davis; T T McMahon; J Schornack; K Zadnik
Journal:  Optom Vis Sci       Date:  1998-04       Impact factor: 1.973

4.  Analysis of Intrasession Repeatability of Ocular Aberrometric Measurements and Validation of Keratometry Provided by a New Integrated System in Mild to Moderate Keratoconus.

Authors:  David P Piñero; Roberto Soto-Negro; Pedro Ruiz-Fortes; Rafael J Pérez-Cambrodí; Hideki Fukumitsu
Journal:  Cornea       Date:  2019-09       Impact factor: 2.651

5.  Repeatability and Reproducibility of Virtual Subjective Refraction.

Authors:  Sara Perches; M Victoria Collados; Jorge Ares
Journal:  Optom Vis Sci       Date:  2016-10       Impact factor: 1.973

6.  Clinical Accuracy of the Nidek ARK-1 Autorefractor.

Authors:  Nabin Paudel; Sameep Adhikari; Ajit Thakur; Bhairaja Shrestha; James Loughman
Journal:  Optom Vis Sci       Date:  2019-06       Impact factor: 1.973

7.  Repeatability of refraction and corrected visual acuity in keratoconus. The CLEK Study Group. Collaborative Longitudinal Evaluation of Keratoconus.

Authors:  L J Davis; K B Schechtman; C G Begley; J A Shin; K Zadnik
Journal:  Optom Vis Sci       Date:  1998-12       Impact factor: 1.973

8.  Spherical subjective refraction with a novel 3D virtual reality based system.

Authors:  Jaume Pujol; Juan Carlos Ondategui-Parra; Llorenç Badiella; Carles Otero; Meritxell Vilaseca; Mikel Aldaba
Journal:  J Optom       Date:  2016-02-05

9.  Validation of refraction and anterior segment parameters by a new multi-diagnostic platform (VX120).

Authors:  Ariela Gordon-Shaag; David P Piñero; Cyril Kahloun; David Markov; Tzadok Parnes; Liat Gantz; Einat Shneor
Journal:  J Optom       Date:  2018-03-09

10.  Clinical evaluation of an automated subjective refraction method implemented in a computer-controlled motorized phoropter.

Authors:  Carles Otero; Mikel Aldaba; Jaume Pujol
Journal:  J Optom       Date:  2018-10-30
View more
  1 in total

Review 1.  Dioptric power and refractive behaviour: a review of methods and applications.

Authors:  Alan Rubin; Tanya Evans; Nabeela Hasrod
Journal:  BMJ Open Ophthalmol       Date:  2022-04-01