Literature DB >> 19921769

Effect of a suction device for femtosecond laser on anterior chamber depth and crystalline lens position measured by OCT.

Kathleen S Kunert1, Marcus Blum, Matthias Reich, Manfred Dick, Christoph Russmann.   

Abstract

PURPOSE: To investigate the position and stability of the crystalline lens after application of a suction device containing a contact lens and a vacuum unit for the treatment of presbyopia using a femtosecond laser.
METHODS: Twenty presbyopic (44.4+/-4.3 years) and 5 pre-presbyopic patients (31.6+/-3.8 years) were included. The anterior chamber depth, along with the position of the lens, was investigated before and after application of the suction device with optical coherence tomography (Visante OCT; Carl Zeiss Meditec AG). The type of suction device is routinely used for femtosecond LASIK with the VisuMax laser (Carl Zeiss Meditec AG).
RESULTS: In both groups, there was a reduction in anterior chamber depth of approximately 700 microm due to the suction device, and the anterior chamber depth achieved was stable. The maximum variation was 160 microm. At the periphery of the crystalline lens, there were movements up to 310 microm axially and 470 microm laterally.
CONCLUSIONS: The study proves that once the suction device has been applied, the crystalline lens is stable enough to undergo presbyopic laser therapy. However, the reduction in anterior chamber depth induced by the suction device showed significant individual variation. The exact position of the lens should therefore be measured immediately before laser surgery. Copyright 2009, SLACK Incorporated.

Entities:  

Mesh:

Year:  2009        PMID: 19921769     DOI: 10.3928/1081597X-20091016-06

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  3 in total

1.  [Femtosecond lenticule extraction (FLEx)].

Authors:  M Blum; W Sekundo
Journal:  Ophthalmologe       Date:  2010-10       Impact factor: 1.059

2.  Femtosecond laser treatment of the crystalline lens: a 1-year study of possible cataractogenesis in minipigs.

Authors:  Roland Ackermann; Kathleen S Kunert; Robert Kammel; Sabine Bischoff; Stephanie C Bühren; Harald Schubert; Marcus Blum; Stefan Nolte
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-08-19       Impact factor: 3.117

3.  Short-term changes in the anterior segment and retina after small incision lenticule extraction.

Authors:  Yanwei Chen; Huaping Liao; Yue Sun; Xi Shen
Journal:  BMC Ophthalmol       Date:  2020-10-07       Impact factor: 2.209