The C-MAC videolaryngoscope: first experiences with a new device for videolaryngoscopy-guided intubation.

Erol Cavus1, Joerg Kieckhaefer, Volker Doerges, Thora Moeller, Carsten Thee, Klaus Wagner.   

Abstract

We studied the efficacy of the C-MAC (Karl Storz, Tuttlingen, Germany), a new portable videolaryngoscope, in 60 patients during routine induction of anesthesia. It was possible to insert the blade (Size 3) of the C-MAC and to get a view of the glottis on the first attempt in all patients. Tracheal intubation also was successful in all 60 patients; 52 patients were intubated on the first attempt, 6 on the second, and 2 on the third. In 8 patients (13%), a gum elastic bougie guide was required. A Cormack-Lehane Class 1 view of the glottis was seen in 30 patients without external manipulation and in 45 with external manipulation, Class 2a view in 22 without and 12 with, Class 2b in 2 without, and Class 3 in 2 without manipulation. The median time taken for tracheal intubation was 16 s (range, 6-58 s). In addition, we describe our experience with 3 patients in whom we had unexpected difficulty with direct laryngoscopy with a conventional Macintosh laryngoscope (Cormack-Lehane Class 3, 4, and 4, respectively). These patients' airways were successfully managed on the first attempt when using the C-MAC Size 4 blade (improvement to Cormack-Lehane Class 1, 2a, and 2b, respectively) in a modified manner by uploading the epiglottis, which is known as "straight blade technique."

Entities:  

Mesh:

Year:  2009        PMID: 19917623     DOI: 10.1213/ANE.0b013e3181c5bce5

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  36 in total

1.  Hemodynamic responses to endotracheal intubation performed with video and direct laryngoscopy in patients scheduled for major cardiac surgery.

Authors:  Gamze Sarkılar; Mehmet Sargın; Tuba Berra Sarıtaş; Hale Borazan; Funda Gök; Alper Kılıçaslan; Şeref Otelcioğlu
Journal:  Int J Clin Exp Med       Date:  2015-07-15

2.  [Implementation of the laryngeal tube for prehospital airway management: training of 1,069 emergency physicians and paramedics].

Authors:  R Schalk; T Auhuber; O Haller; L Latasch; S Wetzel; C F Weber; M Ruesseler; C Byhahn
Journal:  Anaesthesist       Date:  2012-01-25       Impact factor: 1.041

3.  Implementation of a combo videolaryngoscope for intubation in critically ill patients: a before-after comparative study.

Authors:  Audrey De Jong; Noémie Clavieras; Matthieu Conseil; Yannael Coisel; Pierre-Henri Moury; Yvan Pouzeratte; Moussa Cisse; Fouad Belafia; Boris Jung; Gérald Chanques; Nicolas Molinari; Samir Jaber
Journal:  Intensive Care Med       Date:  2013-09-18       Impact factor: 17.440

4.  Cuff inflation to aid nasotracheal intubation using the C-MAC videolaryngoscope.

Authors:  H K Baddoo; B J Phillips
Journal:  Ghana Med J       Date:  2011-06

Review 5.  [Special aspects of the emergency medical care of head and neck injuries].

Authors:  M Helm; H Maier; C Stork; L Lampl; M Bernhard
Journal:  HNO       Date:  2011-08       Impact factor: 1.284

6.  In reply: advantage of Parker Flex-tip Tube(®) in endotracheal intubation using AirwayScope(®) videolaryngoscope.

Authors:  Akihiro Suzuki; Akihito Tampo; Takayuki Kunisawa
Journal:  J Anesth       Date:  2012-10-19       Impact factor: 2.078

7.  [Out-of-hospital airway management in trauma patients : Experiences with the C-MAC® video laryngoscope].

Authors:  B Hossfeld; A Jongebloed; L Lampl; M Helm
Journal:  Unfallchirurg       Date:  2016-06       Impact factor: 1.000

8.  Management of the Difficult Paediatric Airway with a Simple Fiberoptic-Assisted Laryngoscope: A Report of Two Cases with Pierre Robin and Patau's (Trisomy 13) Syndrome.

Authors:  Alper Kılıçaslan; Atilla Erol; Ahmet Topal; Tayfun Et; Şeref Otelcioğlu
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-07-11

9.  Comparison of the C-MAC D-Blade, Conventional C-MAC, and Macintosh Laryngoscopes in Simulated Easy and Difficult Airways.

Authors:  Alper Kılıçaslan; Ahmet Topal; Atilla Erol; Sema Tuncer Uzun
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-08-01

10.  [Reintubation using the C-MAC videolaryngoscope. Implementation in patients with difficult airways initially managed with in situ laryngeal tubes].

Authors:  R Schalk; C F Weber; C Byhahn; C Reyher; D Stay; K Zacharowski; D Meininger
Journal:  Anaesthesist       Date:  2012-08-29       Impact factor: 1.041