[Performance evaluation of the Piccolo xpress Point-of-care Chemistry Analyzer].

Hyunwoong Park1, Dae Hyun Ko, Jin Q Kim, Sang Hoon Song.   

Abstract

BACKGROUND: Point-of-care (POC) tests are used increasingly due to fast results and simple test procedures, which enables rapid diagnosis and therapeutic monitoring. We evaluated the performance of the Piccolo xpress Chemistry Analyzer (Abaxis, USA) a POC chemistry analyzer.
METHODS: Fourteen analytes, Na(+), K(+), Cl(-), Ca(2+), total carbon dioxide, AST, ALT, total bilirubin, alkaline phosphatase, blood urea nitrogen, creatinine, albumin, total protein, and glucose; were measured simultaneously with a 100 microL of whole blood sample using a Comprehensive Metabolic Reagent disk. Within-run and total precision and linearity were evaluated according to CLSI EP15-A and EP6-A guidelines, respectively. Comparison with a central laboratory chemistry analyzer was performed using 144 patient samples.
RESULTS: The coefficients of variations of within-run and total precision were all within 5% for three levels except for total carbon dioxide, ALT, alkaline phosphatase, total bilirubin, and creatinine in low level, and creatinine in middle level. The results of 14 analytes were linear within a commonly encountered range in clinical samples (r(2)> or =0.98). More than 10% of samples in Na(+), AST, ALT, glucose, BUN did not satisfy CLIA analytical quality requirement.
CONCLUSIONS: The Piccolo xpress Chemistry Analyzer can analyze multiple analytes with a minimal amount of whole blood in a short time. It showed an acceptable performance for precision, linearity and comparison with central laboratory analyzer. It can be useful as a screening tests modality in mobile clinics, ambulances, and field clinics for military use, and for pediatric patients from whom enough sample volume is difficult to obtain.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19893352     DOI: 10.3343/kjlm.2009.29.5.430

Source DB:  PubMed          Journal:  Korean J Lab Med        ISSN: 1598-6535


  4 in total

1.  Evaluation of the LABGEO PT10 Point-Of-Care Testing Device: Comparison of Analyte Measurements in Capillary Whole Blood and Lithium Heparin Whole Blood Samples With Those in Central Laboratory.

Authors:  Jinsook Lim; Hyunjin Kim; Sun Hoe Koo; Gye Cheol Kwon
Journal:  J Clin Lab Anal       Date:  2016-09-17       Impact factor: 2.352

2.  Point-of-care creatinine tests to assess kidney function for outpatients requiring contrast-enhanced CT imaging: systematic reviews and economic evaluation.

Authors:  Mark Corbett; Ana Duarte; Alexis Llewellyn; James Altunkaya; Melissa Harden; Martine Harris; Simon Walker; Stephen Palmer; Sofia Dias; Marta Soares
Journal:  Health Technol Assess       Date:  2020-08       Impact factor: 4.014

3.  Assessment of liver function tests on Piccolo Xpress point of care chemistry analyzer in a pediatric hospital.

Authors:  Neval Akbas; Graciela Gonzalez; Rachel Edwards; Sridevi Devaraj
Journal:  Pract Lab Med       Date:  2015-10-03

4.  Novel Opportunities for Improving the Quality of Preanalytical Phase. A Glimpse to the Future?

Authors:  Giuseppe Lippi; Janne Cadamuro
Journal:  J Med Biochem       Date:  2017-10-28       Impact factor: 3.402

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.