Importance of cross-calibration when replacing DXA scanners: QDR4500W and Discovery Wi.

Margaret K Covey1, Donald L Smith, Jean K Berry, Eileen D Hacker.   

Abstract

The aim of the study was to determine reproducibility and extent of agreement between 2 dual-energy X-ray absorptiometers (Hologic QDR4500W, Discovery Wi). The average age of the sample (n = 42) was 50.4 (SD = 9.9) years old and 27.1 (SD = 6.1) kg/m2 body mass index. Four scans were performed with each subject (2 on each device) over approximately 3 weeks. Whole body, proximal femur, and spine scans were performed at each visit. Major variables were whole body bone mineral content (BMC), fat mass, and nonosseous lean mass, and bone mineral density (BMD) of total proximal femur, femoral neck, total spine. Bland and Altman plots assessed the extent of the agreement. Regression analysis was used to develop correction equations if indicated. Both devices demonstrated good precision for whole body composition and BMD of central sites (< 1% different). Interdevice agreement was acceptable for BMD of central sites (< 1% different), but there were systematic differences for whole body composition between the 2 devices. It was concluded that when replacing an existing scanner with a new model, in vivo cross-calibration is important to ensure comparability of scan data, especially for whole body composition.

Mesh:

Year:  2008        PMID: 19886469     DOI: 10.1891/1061-3749.16.3.155

Source DB:  PubMed          Journal:  J Nurs Meas        ISSN: 1061-3749


  5 in total

1.  A novel injectable porous surface modified bioactive bone cement for vertebroplasty: an in vivo biomechanical and osteogenic study in a rabbit osteoporosis model.

Authors:  Jun Chen; Jin Yu; Qiang He; Xiong Zhao; Hongxun Sang; Wei Lei; Zixiang Wu; Jingyuan Chen
Journal:  Am J Transl Res       Date:  2015-03-15       Impact factor: 4.060

2.  Association of circulating dipeptidyl-peptidase 4 levels with osteoporotic fracture in postmenopausal women.

Authors:  H Kim; K H Baek; S-Y Lee; S H Ahn; S H Lee; J-M Koh; Y Rhee; C H Kim; D-Y Kim; M-I Kang; B-J Kim; Y-K Min
Journal:  Osteoporos Int       Date:  2016-11-19       Impact factor: 4.507

3.  Effect of supervised and home exercise training on bone mineral density among breast cancer patients. A 12-month randomised controlled trial.

Authors:  T Saarto; H Sievänen; P Kellokumpu-Lehtinen; R Nikander; L Vehmanen; R Huovinen; H Kautiainen; S Järvenpää; H M Penttinen; M Utriainen; A S Jääskeläinen; A Elme; J Ruohola; T Palva; H Vertio; M Rautalahti; M Fogelholm; R Luoto; C Blomqvist
Journal:  Osteoporos Int       Date:  2011-09-03       Impact factor: 4.507

4.  Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual-Energy X-Ray Densitometers for Bone Mineral Measurements.

Authors:  J Saarelainen; M Hakulinen; T Rikkonen; H Kröger; M Tuppurainen; H Koivumaa-Honkanen; R Honkanen; M Hujo; J S Jurvelin
Journal:  J Osteoporos       Date:  2016-04-27

5.  Investigating the level of agreement of two positioning protocols when using dual energy X-ray absorptiometry in the assessment of body composition.

Authors:  Flinn Shiel; Carl Persson; Vini Simas; James Furness; Mike Climstein; Ben Schram
Journal:  PeerJ       Date:  2017-10-16       Impact factor: 2.984

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.