Continuous noninvasive glucose monitoring technology based on "occlusion spectroscopy".

Orna Amir1, Daphna Weinstein, Silviu Zilberman, Malka Less, Daniele Perl-Treves, Harel Primack, Aharon Weinstein, Efi Gabis, Boris Fikhte, Avraham Karasik.   

Abstract

BACKGROUND: A truly noninvasive glucose-sensing device could revolutionalize diabetes treatment by leading to improved compliance with recommended glucose levels, thus reducing the long-term complications and cost of diabetes. Herein, we present the technology and evaluate the efficacy of a truly noninvasive device for continuous blood glucose monitoring, the NBM (OrSense Ltd.).
METHODS: In vitro analysis was used to validate the technology and algorithms. A clinical study was performed to quantify the in vivo performance of the NBM device. A total of 23 patients with type 1 (n = 12) and type 2 (n = 11) diabetes were enrolled in the clinical study and participated in 111 sessions. Accuracy was assessed by comparing NBM data with paired self-monitoring of blood glucose meter readings.
RESULTS: In vitro experiments showed a strong correlation between calculated and actual glucose concentrations. The clinical trial produced a total of 1690 paired glucose values (NBM vs reference). In the paired data set, the reference glucose range was 40-496 mg/dl. No systematic bias was found at any of the glucose levels examined (70, 100, 150, and 200 mg/dl). The mean relative absolute difference was 17.2%, and a Clarke error grid analysis showed that 95.5% of the measurements fall within the clinically acceptable A&B regions (zone A, 69.7%; and zone B, 25.7%).
CONCLUSIONS: This study indicates the potential use of OrSense's NBM device as a noninvasive sensor for continuous blood glucose evaluation. The device was safe and well tolerated.

Entities:  

Keywords:  noninvasive glucose monitoring; occlusion spectroscopy; self-monitoring of blood glucose

Year:  2007        PMID: 19885108      PMCID: PMC2769638          DOI: 10.1177/193229680700100403

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  18 in total

Review 1.  Non-invasive glucose measurement technologies: an update from 1999 to the dawn of the new millennium.

Authors:  Omar S Khalil
Journal:  Diabetes Technol Ther       Date:  2004-10       Impact factor: 6.118

Review 2.  Continuous glucose monitoring: roadmap for 21st century diabetes therapy.

Authors:  David C Klonoff
Journal:  Diabetes Care       Date:  2005-05       Impact factor: 19.112

3.  Measurement of circulating glucose concentrations: the time is now for consistency among methods and types of samples.

Authors:  Michael W Steffes; David B Sacks
Journal:  Clin Chem       Date:  2005-09       Impact factor: 8.327

4.  Standards of medical care in diabetes--2006.

Authors: 
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

5.  Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial.

Authors:  Satish Garg; Howard Zisser; Sherwyn Schwartz; Timothy Bailey; Roy Kaplan; Samuel Ellis; Lois Jovanovic
Journal:  Diabetes Care       Date:  2006-01       Impact factor: 19.112

Review 6.  Disease management in the young diabetic patient: glucose monitoring, coping skills, and treatment strategies.

Authors:  Stuart A Weinzimer; Elizabeth A Doyle; William V Tamborlane
Journal:  Clin Pediatr (Phila)       Date:  2005-06       Impact factor: 1.168

7.  The influence of glucose concentration upon the transport of light in tissue-simulating phantoms.

Authors:  M Kohl; M Essenpreis; M Cope
Journal:  Phys Med Biol       Date:  1995-07       Impact factor: 3.609

8.  Noninvasive continuous glucose monitoring using photoacoustic technology-results from the first 62 subjects.

Authors:  Ram Weiss; Yevgeny Yegorchikov; Alexander Shusterman; Itamar Raz
Journal:  Diabetes Technol Ther       Date:  2007-02       Impact factor: 6.118

9.  Clinical performance of CGMS in type 1 diabetic patients treated by continuous subcutaneous insulin infusion using insulin analogs.

Authors:  Bruno Guerci; Michèle Floriot; Philip Böhme; Danielle Durain; Muriel Benichou; Stéphanie Jellimann; Pierre Drouin
Journal:  Diabetes Care       Date:  2003-03       Impact factor: 19.112

Review 10.  An overview of minimally invasive technologies.

Authors:  B H Ginsberg
Journal:  Clin Chem       Date:  1992-09       Impact factor: 8.327

View more
  8 in total

Review 1.  Noninvasive Blood Glucose Monitoring Systems Using Near-Infrared Technology-A Review.

Authors:  Aminah Hina; Wala Saadeh
Journal:  Sensors (Basel)       Date:  2022-06-27       Impact factor: 3.847

2.  Noninvasive, wearable, and tunable electromagnetic multisensing system for continuous glucose monitoring, mimicking vasculature anatomy.

Authors:  Jessica Hanna; Moussa Bteich; Youssef Tawk; Ali H Ramadan; Batoul Dia; Fatima A Asadallah; Aline Eid; Rouwaida Kanj; Joseph Costantine; Assaad A Eid
Journal:  Sci Adv       Date:  2020-06-10       Impact factor: 14.136

Review 3.  Review of Non-invasive Glucose Sensing Techniques: Optical, Electrical and Breath Acetone.

Authors:  Maryamsadat Shokrekhodaei; Stella Quinones
Journal:  Sensors (Basel)       Date:  2020-02-25       Impact factor: 3.576

4.  Analytical Model for Blood Glucose Detection Using Electrical Impedance Spectroscopy.

Authors:  Bruna Gabriela Pedro; David William Cordeiro Marcôndes; Pedro Bertemes-Filho
Journal:  Sensors (Basel)       Date:  2020-12-04       Impact factor: 3.576

5.  Development of a Noninvasive Blood Glucose Monitoring System Prototype: Pilot Study.

Authors:  Maria Valero; Priyanka Pola; Oluwaseyi Falaiye; Katherine H Ingram; Liang Zhao; Hossain Shahriar; Sheikh Iqbal Ahamed
Journal:  JMIR Form Res       Date:  2022-08-26

Review 6.  Italian contributions to the development of continuous glucose monitoring sensors for diabetes management.

Authors:  Giovanni Sparacino; Mattia Zanon; Andrea Facchinetti; Chiara Zecchin; Alberto Maran; Claudio Cobelli
Journal:  Sensors (Basel)       Date:  2012-10-12       Impact factor: 3.576

7.  Design and In Vitro Interference Test of Microwave Noninvasive Blood Glucose Monitoring Sensor.

Authors:  Heungjae Choi; Jack Naylon; Steve Luzio; Jan Beutler; James Birchall; Chris Martin; Adrian Porch
Journal:  IEEE Trans Microw Theory Tech       Date:  2015-10-01       Impact factor: 3.599

8.  Low-cost portable microwave sensor for non-invasive monitoring of blood glucose level: novel design utilizing a four-cell CSRR hexagonal configuration.

Authors:  Ala Eldin Omer; George Shaker; Safieddin Safavi-Naeini; Hamid Kokabi; Georges Alquié; Frédérique Deshours; Raed M Shubair
Journal:  Sci Rep       Date:  2020-09-16       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.