Comparison of IOP measurement by ocular response analyzer, dynamic contour, Goldmann applanation, and noncontact tonometry.

Banu Oncel1, Umut Dinc, Faruk Orge, Belkis Yalvac.   

Abstract

PURPOSE: To assess the agreement among ocular response analyzer (ORA), dynamic contour tonometry (DCT), Goldmann applanation tonometry (GAT), and noncontact tonometry (NCT), and to determine the effects of central corneal thickness (CCT) and corneal hysteresis (CH) on intraocular pressure (IOP) measurements with these devices.
METHODS: Sixty healthy volunteers were recruited. The average of ORA (corneal compensated IOP [IOP-ORAcc] and Goldmann-correlated IOP [IOP-ORAg]), DCT, GAT, and NCT levels were compared and the devices were examined with respect to CCT and CH. One-way analysis of variance and Pearson correlation tests were used for statistical analysis.
RESULTS: Mean CCT was 543.28+/-36.2 mum and mean CH was 10.87+/-1.27 mmHg. Mean IOP obtained using DCT was 16.4+/-2.9 mmHg, whereas those provided by ORA were 15.8+/-2.9 mmHg for IOPORAcc and 15.1+/-2.8 mmHg for IOP-ORAg. The mean IOPs obtained using GAT and NCT were 14.8+/-3.1 mmHg and 14.7+/-2.8 mmHg. The differences between the measurements of DCT and GAT, NCT, and IOP-ORAg were statistically significant (p=0.002, p=0.002, and p=0.015, respectively), but not between DCT and IOP-ORAcc (p=0.263). We did not detect a significant correlation between DCT and CCT, and IOP-ORAcc and CCT (p=0.081 and p=0.093). No correlation was observed between IOP-ORAg and CCT (p=0.067). We found a significant correlation between GAT and CCT, and NCT and CCT (p=0.004 and p=0.001). CH was detected to be correlated only to IOP-ORAcc (p=0.017).
CONCLUSIONS: IOP readings detected by DCT and IOP-ORAcc were not found to be clinically interchangeable with GAT and NCT readings. IOP-ORAcc and DCT readings may be regarded as comparable and independent of CCT, in the range of the normal IOP.

Entities:  

Mesh:

Year:  2009        PMID: 19882585     DOI: 10.1177/112067210901900607

Source DB:  PubMed          Journal:  Eur J Ophthalmol        ISSN: 1120-6721            Impact factor:   1.922


  11 in total

Review 1.  The thick and thin of the central corneal thickness in glaucoma.

Authors:  Graham W Belovay; Ivan Goldberg
Journal:  Eye (Lond)       Date:  2018-02-15       Impact factor: 3.775

2.  Repeatability of the Novel Intraocular Pressure Measurement From Corvis ST.

Authors:  Masato Matsuura; Hiroshi Murata; Yuri Fujino; Mieko Yanagisawa; Yoshitaka Nakao; Shunsuke Nakakura; Yoshiaki Kiuchi; Ryo Asaoka
Journal:  Transl Vis Sci Technol       Date:  2019-06-24       Impact factor: 3.283

3.  Ocular surface displacement with and without contact lenses during non-contact tonometry.

Authors:  Ulfah Rimayanti; Yoshiaki Kiuchi; Shohei Uemura; Joji Takenaka; Hideki Mochizuki; Makoto Kaneko
Journal:  PLoS One       Date:  2014-04-29       Impact factor: 3.240

Review 4.  Optical coherence elastography in ophthalmology.

Authors:  Mitchell A Kirby; Ivan Pelivanov; Shaozhen Song; Łukasz Ambrozinski; Soon Joon Yoon; Liang Gao; David Li; Tueng T Shen; Ruikang K Wang; Matthew O'Donnell
Journal:  J Biomed Opt       Date:  2017-12       Impact factor: 3.170

Review 5.  Review of Corneal Biomechanical Properties Following LASIK and SMILE for Myopia and Myopic Astigmatism.

Authors:  Iben Bach Damgaard; Mohamed Reffat; Jesper Hjortdal
Journal:  Open Ophthalmol J       Date:  2018-07-27

6.  Intraocular Pressure Based on Dynamic Bidirectional Applanation and Air-puff Tonometry: A Comparative Study.

Authors:  Ayse E Bahadir Kilavuzoglu; Cemile B Cosar; Ali Rc Celebi; Ugur E Al Parmak
Journal:  J Curr Glaucoma Pract       Date:  2019 May-Aug

Review 7.  Advances in Biomechanical Parameters for Screening of Refractive Surgery Candidates: A Review of the Literature, Part III.

Authors:  Majid Moshirfar; Mahsaw N Motlagh; Michael S Murri; Hamed Momeni-Moghaddam; Yasmyne C Ronquillo; Phillip C Hoopes
Journal:  Med Hypothesis Discov Innov Ophthalmol       Date:  2019

8.  Changes of intraocular pressure and refractive status in children following cycloplegic refraction with 1% cyclopentolate and 1% tropicamide.

Authors:  Kuo-Chi Hung; Hsiu-Mei Huang; Pei-Wen Lin
Journal:  Taiwan J Ophthalmol       Date:  2015-08-24

9.  Comparison of different intraocular pressure measurement techniques in normal eyes and post small incision lenticule extraction.

Authors:  Mohamed Hosny; Fayrouz Aboalazayem; Hoda El Shiwy; Mohsen Salem
Journal:  Clin Ophthalmol       Date:  2017-07-14

10.  Corneal Biomechanical Properties of Keratoconic Eyes Following Penetrating Keratoplasty.

Authors:  Hamidu Gobeka; Özlem Barut Selver; Melis Palamar Onay; Sait Eğrilmez; Ayşe Yağcı
Journal:  Turk J Ophthalmol       Date:  2018-09-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.