We investigated the efficacy of combining highly cross-linked polyethylene with ceramic heads on further reduction in polyethylene wear compared with the combination with cobalt-chromium heads via PolyWare computer-assisted method. A prospective cohort study was performed on 102 cementless total hip arthroplasties using Longevity (Zimmer, Warsaw, Ind) highly cross-linked polyethylene liners. Either 26-mm zirconia heads or 26-mm cobalt-chromium heads were randomly used in 51 hips each. At a mean follow-up of 6.7 years, no significant differences were identified between the groups for total penetration rate and steady-state wear rate. Osteolysis was not observed in any hips in either group. In conclusion, no advantage was seen for the 26-mm zirconia head compared with the 26-mm cobalt-chromium head in this period.
Copyright © 2010 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19879725     DOI: 10.1016/j.arth.2009.09.006

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  8 in total

1.  Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  George K Karidakis; Theofilos Karachalios
Journal:  Clin Orthop Relat Res       Date:  2015-12       Impact factor: 4.176

2.  Long-term outcomes of total hip arthroplasty in patients younger than 55 years: a systematic review of the contemporary literature

Authors:  Xin Yu Mei; Ying Jia Gong; Oleg Safir; Allan Gross; Paul Kuzyk
Journal:  Can J Surg       Date:  2019-08-01       Impact factor: 2.089

3.  Reduced wear of cross-linked UHMWPE using magnesia-stabilized zirconia femoral heads in a hip simulator.

Authors:  Marcel E Roy; Leo A Whiteside; Mark E Magill; Brian J Katerberg
Journal:  Clin Orthop Relat Res       Date:  2011-02-11       Impact factor: 4.176

Review 4.  History and systematic review of wear and osteolysis outcomes for first-generation highly crosslinked polyethylene.

Authors:  Steven M Kurtz; Heather A Gawel; Jasmine D Patel
Journal:  Clin Orthop Relat Res       Date:  2011-08       Impact factor: 4.176

5.  Charnley-Kerboull total hip arthroplasty combining zirconia on polyethylene. A minimum eight-year follow-up prospective study.

Authors:  Jean Langlois; Samer El Hage; Faouzi Madi; Jean-Pierre Courpied; Marcel Kerboull; Moussa Hamadouche
Journal:  Int Orthop       Date:  2013-01-31       Impact factor: 3.075

6.  Is there any difference in survivorship of total hip arthroplasty with different bearing surfaces? A systematic review and network meta-analysis.

Authors:  Si Yin; Dangfeng Zhang; Hui Du; Heng Du; Zhanhai Yin; Yusheng Qiu
Journal:  Int J Clin Exp Med       Date:  2015-11-15

7.  In vivo wear performance of highly cross-linked polyethylene vs. yttria stabilized zirconia and alumina stabilized zirconia at a mean seven-year follow-up.

Authors:  Masahiro Hasegawa; Akihiro Sudo
Journal:  BMC Musculoskelet Disord       Date:  2013-05-01       Impact factor: 2.362

8.  Comparison of conventional polyethylene wear and signs of cup failure in two similar total hip designs.

Authors:  Thomas B Pace; Kevin C Keith; Estefania Alvarez; Rebecca G Snider; Stephanie L Tanner; John D Desjardins
Journal:  Adv Orthop       Date:  2013-04-11