Risk of extravasation after power injection of contrast media via the proximal port of multilumen central venous catheters: case report and review of the literature.

C Schummer1, Y Sakr, J Steenbeck, M Gugel, K Reinhart, W Schummer.   

Abstract

PURPOSE: Multilumen central venous catheters (CVCs) are not commonly used for power injection. However, in critically ill patients, CVCs-- most of which do not have FDA approval for power injection--may be the only available venous access.
MATERIALS AND METHODS: The pitfalls of multilumen CVCs are illustrated by a case report of a patient in whom extravasation of intravenously administered contrast medium occurred after power injection in a triple-lumen CVC using the lumen with the port furthest from the catheter tip.
RESULTS: The underlying mechanisms for the displacement of the initially correctly placed right subclavian CVC could include elevation of both arms of the obese patient or the power injection itself. The distances between port openings and catheter tips of various commercially available multilumen CVCs are assessed. We examine the possible caveats of ECG-guided CVC placement for optimal tip position, discuss technical difficulties related to power injection via CVCs, and review commonly used drugs that may cause extravasation injury.
CONCLUSION: Knowledge of the distances between CVC port openings and the catheter tip are essential for safe intravasal administration of fluids. (c) Georg Thieme Verlag KG Stuttgart-New York.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19859861     DOI: 10.1055/s-0028-1109742

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  3 in total

Review 1.  The use of central venous catheters for intravenous contrast injection for CT examinations.

Authors:  A A O Plumb; G Murphy
Journal:  Br J Radiol       Date:  2011-03       Impact factor: 3.039

2.  Comparison of the Quality of Life of Patients with Breast or Colon Cancer with an Arm Vein Port (TIVAD) Versus a Peripherally Inserted Central Catheter (PICC).

Authors:  Brent Burbridge; Hyun Lim; Lynn Dwernychuk; Ha Le; Tehmina Asif; Amer Sami; Shahid Ahmed
Journal:  Curr Oncol       Date:  2021-04-09       Impact factor: 3.677

3.  Power port contrast medium flushing and trapping: impact of temperature, an in vitro experimental study.

Authors:  Gérard Guiffant; Jean Jacques Durussel; Patrice Flaud; Laurent Royon; Pierre Yves Marcy; Jacques Merckx
Journal:  Med Devices (Auckl)       Date:  2013-09-03
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.