Diffusion-weighted MR imaging of the liver at 3.0 Tesla using TRacking Only Navigator echo (TRON): a feasibility study.

Marko K Ivancevic1, Thomas C Kwee, Taro Takahara, Tetsuo Ogino, Hero K Hussain, Peter S Liu, Thomas L Chenevert.   

Abstract

PURPOSE: To assess the feasibility of TRacking Only Navigator echo (TRON) for diffusion-weighted magnetic resonance imaging (DWI) of the liver at 3.0T.
MATERIALS AND METHODS: Ten volunteers underwent TRON, respiratory triggered, and free breathing DWI of the liver at 3.0 Tesla (T). Scan times were measured. Image sharpness, degree of stair-step and stripe artifacts for the three methods were assessed by two observers.
RESULTS: Mean scan times of TRON and respiratory triggered DWI relative to free breathing DWI were 34% and 145% longer respectively. In four of eight comparisons (two observers, two b-values, two slice orientations), TRON DWI image sharpness was significantly better than free breathing DWI, but inferior to respiratory triggered DWI. In two of four comparisons (two observers, two b-values), degree of stair-step artifacts in TRON DWI was significantly lower than in respiratory triggered DWI. Degree of stripe artifacts between the three methods was not significantly different.
CONCLUSION: DWI of the liver at 3.0T using TRON is feasible. Image sharpness in TRON DWI is superior to that in free breathing DWI. Although image sharpness of respiratory triggered DWI is still better, TRON DWI requires less scan time and reduces stair-step artifacts.

Mesh:

Year:  2009        PMID: 19856431     DOI: 10.1002/jmri.21939

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  5 in total

Review 1.  Diffusion weighted imaging in the liver.

Authors:  Petra G Kele; Eric J van der Jagt
Journal:  World J Gastroenterol       Date:  2010-04-07       Impact factor: 5.742

Review 2.  Predicting and monitoring cancer treatment response with diffusion-weighted MRI.

Authors:  Harriet C Thoeny; Brian D Ross
Journal:  J Magn Reson Imaging       Date:  2010-07       Impact factor: 4.813

3.  Quantitative liver ADC measurements using diffusion-weighted MRI at 3 Tesla: evaluation of reproducibility and perfusion dependence using different techniques for respiratory compensation.

Authors:  Nis Elbrønd Larsen; Søren Haack; Lars Peter Skovgaard Larsen; Erik Morre Pedersen
Journal:  MAGMA       Date:  2013-03-13       Impact factor: 2.310

4.  The Accuracy of ADC Measurements in Liver Is Improved by a Tailored and Computationally Efficient Local-Rigid Registration Algorithm.

Authors:  Hossein Ragheb; Neil A Thacker; Jean-Marie Guyader; Stefan Klein; Nandita M deSouza; Alan Jackson
Journal:  PLoS One       Date:  2015-07-23       Impact factor: 3.240

5.  ESR statement on the stepwise development of imaging biomarkers.

Authors: 
Journal:  Insights Imaging       Date:  2013-02-09
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.