Intraocular pressure in keratoconus.

Scott A Read1, Michael J Collins.   

Abstract

PURPOSE: This study aimed to investigate intraocular pressure (IOP) measurements taken using the dynamic contour tonometer (DCT) and the non-contact tonometer (NCT) in subjects with keratoconus.
METHODS: Intraocular pressure was measured in 20 keratoconus subjects and 20 age-matched control subjects using the DCT and NCT instruments. Central and off-centre measures were taken with the DCT in order to highlight any systematic errors associated with corneal biomechanical factors. Measures of anterior and posterior corneal topography and thickness were also taken in each subject.
RESULTS: No significant difference was found between the central and off-centre DCT IOP readings for the keratoconus subjects and age-matched controls (p > 0.05). The average DCT IOP was 14.2 ± 1.4 mmHg in the keratoconus subjects and 14.2 ± 1.6 mmHg in the controls. However, the average NCT readings differed significantly (p < 0.001) between the keratoconus subjects (9.2 ± 1.5 mmHg) and the controls (12.9 ± 2.4 mmHg). Intraocular pressure measurements taken with the DCT showed no significant (p > 0.05) correlation with the severity of keratoconus as determined through measures of corneal topography and thickness. Equivalent IOP measurements taken with the NCT correlated significantly with certain measures of corneal curvature and thickness in the keratoconus population. The difference between DCT and NCT IOP was strongly correlated with measures of corneal topography and thickness, with differences increasing in more advanced keratoconus.
CONCLUSIONS: The measurements obtained with the DCT do not appear to be dependent upon corneal factors, unlike those taken with the NCT. The presence or severity of keratoconus was not correlated with DCT IOP values.
© 2009 The Authors. Journal compilation © 2009 Acta Ophthalmol.

Entities:  

Mesh:

Year:  2009        PMID: 19845560     DOI: 10.1111/j.1755-3768.2009.01690.x

Source DB:  PubMed          Journal:  Acta Ophthalmol        ISSN: 1755-375X            Impact factor:   3.761


  4 in total

1.  The influence of corneal geometrical and biomechanical properties on tonometry readings in keratoconic eyes.

Authors:  Mustafa Değer Bilgeç; Eray Atalay; Ömer Sözer; Hüseyin Gürsoy; Muzaffer Bilgin; Nilgün Yıldırım
Journal:  Int Ophthalmol       Date:  2019-12-02       Impact factor: 2.031

2.  Twenty-four hours intraocular pressure in keratoconic eyes assessed by applanation tonometry and Tono-Pen AVIA.

Authors:  Rafael Vidal Merola; Sebastião Cronemberger; Artur William Veloso; Alberto Diniz-Filho
Journal:  Int J Ophthalmol       Date:  2022-01-18       Impact factor: 1.779

3.  Distribution of IOP measured with an air puff tonometer in a young population.

Authors:  Hassan Hashemi; Mehdi Khabazkhoob; Payam Nabovati; Negareh Yazdani; Hadi Ostadimoghaddam; Ehsan Shiralivand; Akbar Derakhshan; AbbasAli Yekta
Journal:  J Curr Ophthalmol       Date:  2017-04-02

4.  Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus.

Authors:  Fatih Özcura; Nilgün Yıldırım; Emre Tambova; Afsun Şahin
Journal:  J Optom       Date:  2016-07-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.