Literature DB >> 1984159

Angiographic follow-up after placement of a self-expanding coronary-artery stent.

P W Serruys1, B H Strauss, K J Beatt, M E Bertrand, J Puel, A F Rickards, B Meier, J J Goy, P Vogt, L Kappenberger.   

Abstract

BACKGROUND: The placement of stents in coronary arteries after coronary angioplasty has been investigated as a way of treating abrupt coronary-artery occlusion related to the angioplasty and of reducing the late intimal hyperplasia responsible for gradual restenosis of the dilated lesion.
METHODS: From March 1986 to January 1988, we implanted 117 self-expanding, stainless-steel endovascular stents (Wallstent) in the native coronary arteries (94 stents) or saphenous-vein bypass grafts (23 stents) of 105 patients. Angiograms were obtained immediately before and after placement of the stent and at follow-up at least one month later (unless symptoms required angiography sooner). The mortality after one year was 7.6 percent (8 patients). Follow-up angiograms (after a mean [+/- SD] of 5.7 +/- 4.4 months) were obtained in 95 patients with 105 stents and were analyzed quantitatively by a computer-assisted system of cardiovascular angiographic analysis. The 10 patients without follow-up angiograms included 4 who died.
RESULTS: Complete occlusion occurred in 27 stents in 25 patients (24 percent); 21 occlusions were documented within the first 14 days after implantation. Overall, immediately after placement of the stent there was a significant increase in the minimal luminal diameter and a significant decrease in the percentage of the diameter with stenosis (changing from a mean [+/- SD] of 1.88 +/- 0.43 to 2.48 +/- 0.51 mm and from 37 +/- 12 to 21 +/- 10 percent, respectively; P less than 0.0001). Later, however, there was a significant decrease in the minimal luminal diameter and a significant increase in the stenosis of the segment with the stent (1.68 +/- 1.78 mm and 48 +/- 34 percent at follow-up). Significant restenosis, as indicated by a reduction of 0.72 mm in the minimal luminal diameter or by an increase in the percentage of stenosis to greater than or equal to 50 percent, occurred in 32 percent and 14 percent of patent stents, respectively.
CONCLUSIONS: Early occlusion remains an important limitation of this coronary-artery stent. Even when the early effects are beneficial, there are frequently late occlusions or restenosis. The place of this form of treatment for coronary artery disease remains to be determined.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1984159     DOI: 10.1056/NEJM199101033240103

Source DB:  PubMed          Journal:  N Engl J Med        ISSN: 0028-4793            Impact factor:   91.245


  69 in total

Review 1.  Clopidogrel with aspirin is the optimal antiplatelet regimen for intracoronary stenting.

Authors:  S R Steinhubl; E J Topol
Journal:  J Thromb Thrombolysis       Date:  1999-06       Impact factor: 2.300

2.  Adjunctive Pharmacologic Treatment: Focus on the Development of Low Molecular Weight Heparins.

Authors: 
Journal:  J Thromb Thrombolysis       Date:  1997       Impact factor: 2.300

Review 3.  Antiplatelet therapy in interventional cardiology: I. Newer oral antiplatelet agents.

Authors:  F H Jafary; C D Kimmelstiel
Journal:  J Thromb Thrombolysis       Date:  2000-02       Impact factor: 2.300

Review 4.  Stent thrombosis: historical perspectives and current trends.

Authors:  D E Cutlip
Journal:  J Thromb Thrombolysis       Date:  2000-08       Impact factor: 2.300

5.  A comparison study of self-expandable stents vs balloon angioplasty alone in femoropopliteal artery occlusions.

Authors:  J Triller; B H Walpoth; P Stirnemann; F Mahler
Journal:  Cardiovasc Intervent Radiol       Date:  1992 Sep-Oct       Impact factor: 2.740

6.  Expandable metal stents for tracheal obstruction: permanent or temporary? A cautionary tale.

Authors:  C R Hind; R J Donnelly
Journal:  Thorax       Date:  1992-09       Impact factor: 9.139

7.  Restenosis after coronary angioplasty: a proposal of new comparative approaches based on quantitative angiography.

Authors:  P W Serruys; D P Foley; P J de Feyter
Journal:  Br Heart J       Date:  1992-10

8.  Electrochemical polishing of 316L stainless steel slotted tube coronary stents.

Authors:  Hui Zhao; Jan Van Humbeeck; Jürgen Sohier; Ivan De Scheerder
Journal:  J Mater Sci Mater Med       Date:  2002-10       Impact factor: 3.896

Review 9.  Robotic-assisted percutaneous coronary intervention--filling an unmet need.

Authors:  Joseph P Carrozza
Journal:  J Cardiovasc Transl Res       Date:  2011-11-09       Impact factor: 4.132

Review 10.  Recent advances in cardiology.

Authors:  C F Shakespeare; D J Coltart
Journal:  Postgrad Med J       Date:  1992-05       Impact factor: 2.401