Yousef Shahin1, Rakesh Choudhary, Saeed Al-Naser, Mark Mullins.
Abstract
We report a case of interposition of a broken trochanteric wire in the hip joint. This caused early wear of the prosthesis and dislocation of the Charnley total hip arthroplasty. The patient was treated with a revision total hip arthroplasty. This rare complication should be taken into consideration when performing a trochanteric osteotomy fixation with wiring in Charnley total hip arthroplasty.Entities:
Year: 2009 PMID: 19829913 PMCID: PMC2740164 DOI: 10.1186/1757-1626-2-7117
Source DB: PubMed Journal: Cases J ISSN: 1757-1626
Figure 1.Post Charnley left total hip replacement plain X-ray of the pelvis.
Figure 2.Plain X-ray of the pelvis showing a dislocated left hip with a broken wire in the left hip.
Figure 3.The acetabular component with a worn bearing surface and the broken wire embedded in it.