Early loosening and secondary dislocation due to a broken trochanteric osteotomy wire following a Charnley total hip arthroplasty: a case report.

Yousef Shahin1, Rakesh Choudhary, Saeed Al-Naser, Mark Mullins.   

Abstract

We report a case of interposition of a broken trochanteric wire in the hip joint. This caused early wear of the prosthesis and dislocation of the Charnley total hip arthroplasty. The patient was treated with a revision total hip arthroplasty. This rare complication should be taken into consideration when performing a trochanteric osteotomy fixation with wiring in Charnley total hip arthroplasty.

Entities:  

Year:  2009        PMID: 19829913      PMCID: PMC2740164          DOI: 10.1186/1757-1626-2-7117

Source DB:  PubMed          Journal:  Cases J        ISSN: 1757-1626


Introduction

Trochanteric osteotomy is a known method used in Charnley total hip arthroplasty. The advantage of the transtrochanteric approach is that it gives a better view of the hip joint. The disadvantages of this method are increased operative time, increased blood loss, trochanteric non union and trochanteric bursitis [1,2]. We report a rare complication associated with trochanteric osteotomy.

Case presentation

A 61 year old British white male was admitted with a dislocation of a left Charnley total hip arthroplasty which was performed in 2005 for a subcapital fracture of the left neck of femur. The operation was successful for three years and the patient was doing well. Immediate post operative plain X-ray of the pelvis was satisfactory (Figure 1). In March 2008 the patient presented with a sudden left hip pain while was getting into bed and was unable to weight bear afterwards. On examination, the left leg was shortened and externally rotated.
Figure 1.

Post Charnley left total hip replacement plain X-ray of the pelvis.

Post Charnley left total hip replacement plain X-ray of the pelvis. Plain X-ray of the pelvis showed a dislocated left Charnley total hip arthroplasty and it also showed a foreign body in the hip joint (broken wire) (Figure 2). The hip was reduced under general anaesthesia. The hip joint was stable up to 90 degrees flexion and 10 degrees of internal rotation.
Figure 2.

Plain X-ray of the pelvis showing a dislocated left hip with a broken wire in the left hip.

Plain X-ray of the pelvis showing a dislocated left hip with a broken wire in the left hip. In April 2008, the patient had a revision of left Charnley total hip arthroplasty with a good result. Intraoperatively, a broken wire was found embedded in the polyethylene acetabular component and the polyethylene bearing surface was found to be worn (Figure 3).
Figure 3.

The acetabular component with a worn bearing surface and the broken wire embedded in it.

The acetabular component with a worn bearing surface and the broken wire embedded in it.

Discussion

Articular interposition of a broken wire in the hip joint is a rare complication following Charnley total hip arthroplasty. Four cases have been reported in the literature. The first case was a 53 year old male who had a revision arthroplasty because of a worn acetabular cup due to an intra-articular wire fragments found two years after Charnley total hip arthroplasty [3]. The second case was a 64 year old male who was found to have a wire on the X-rays after three years of the operation. He did not have any symptoms but after 12 years required revision arthroplasty due to cup wear [4]. The third case was an 80 year old female who experienced multiple dislocations of the hip 10 years following Charnley total hip arthroplasty. The wires were taken out arthroscopically [5]. The fourth case was an 80 year old female who had recurrent dislocations 10 years after the operation. Plain X-rays showed intra-articular wires. She required a revision arthroplasty due to prosthetic components wear [6]. In our case, the patient did not have any problems post operatively for three years until he experienced his first dislocation.

Conclusion

The interposition of broken trochanteric wires in the hip joint post Charnley total hip arthroplasty can cause a third body effect leading to early loosening and early wear of the prosthetic components leading to dislocation of the hip. It is a complication that should be considered when performing a Charnley total hip arthroplasty which might necessitate an early revision surgery.
  5 in total

1.  TRANSPLANTATION OF THE GREATER TROCHANTER IN ARTHROPLASTY OF THE HIP.

Authors:  J CHARNLEY; S FERREIRAADE
Journal:  J Bone Joint Surg Br       Date:  1964-05

2.  Articular interposition of broken trochanteric wires.

Authors:  R R Fakih; R B Treacy
Journal:  Bull Hosp Jt Dis       Date:  1998

3.  Articular interposition of trochanteric wires in a failed total hip replacement.

Authors:  J L Bronson
Journal:  Clin Orthop Relat Res       Date:  1976 Nov-Dec       Impact factor: 4.176

4.  Intra-articular migration of broken trochanteric wires.

Authors:  A B Lim; S Parekh; D N Smith
Journal:  Injury       Date:  1993-03       Impact factor: 2.586

5.  Intraarticular trochanteric wire migration following bilateral total hip arthroplasty. A case report.

Authors:  M A Ritter; J B Meding
Journal:  Orthopedics       Date:  1988-09       Impact factor: 1.390

  5 in total
  1 in total

Review 1.  A systematic review and meta-analysis of complications following the posterior and lateral surgical approaches to total hip arthroplasty.

Authors:  J R Berstock; A W Blom; A D Beswick
Journal:  Ann R Coll Surg Engl       Date:  2015-01       Impact factor: 1.891

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.