Gamma knife output factor measurements using VIP polymer gel dosimetry.

A Moutsatsos1, L Petrokokkinos, P Karaiskos, P Papagiannis, E Georgiou, K Dardoufas, P Sandilos, M Torrens, E Pantelis, I Kantemiris, L Sakelliou, I Seimenis.   

Abstract

PURPOSE: Water equivalent polymer gel dosimeters and magnetic resonance imaging were employed to measure the output factors of the two smallest treatment fields available in a Gamma Knife model C radiosurgery unit, those formed employing the 4 and 8 mm final collimator helmets.
METHODS: Three samples of the VIP normoxic gel formulation were prepared and irradiated so that a single shot of the field whose output factor is to be measured and a single shot of the reference 18 mm field were delivered in each one. Emphasis is given to the development and benchmarking of a refined data processing methodology of reduced uncertainty that fully exploits the 3D dose distributions registered in the dosimeters.
RESULTS: Polymer gel results for the output factor of the 8 mm collimator helmet are found to be in close agreement with the corresponding value recommended by the vendor (0.955 +/- 0.007 versus 0.956, respectively). For the 4 mm collimator helmet, however, polymer gel results suggest an output factor 3% lower than the value recommended by the vendor (0.841 +/- 0.009 versus 0.870, respectively).
CONCLUSIONS: A comparison with corresponding measurements published in the literature indicates that output factor results of this work are in agreement with those obtained using dosimetric systems which, besides fine spatial resolution and lack of angular and dose rate dependence of the dosimeter's response, share with polymer gels the favorable characteristic of minimal radiation field perturbation.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19810502     DOI: 10.1118/1.3183500

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  6 in total

1.  PRESAGE 3D dosimetry accurately measures Gamma Knife output factors.

Authors:  Slade J Klawikowski; James N Yang; John Adamovics; Geoffrey S Ibbott
Journal:  Phys Med Biol       Date:  2014-11-04       Impact factor: 3.609

2.  MRI-based polymer gel dosimetry for validating plans with multiple matrices in Gamma Knife stereotactic radiosurgery.

Authors:  N Gopishankar; Yoichi Watanabe; Vivekanandhan Subbiah
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

3.  Target localization accuracy in frame-based stereotactic radiosurgery: Comparison between MR-only and MR/CT co-registration approaches.

Authors:  Eleftherios P Pappas; Ioannis Seimenis; Panagiotis Kouris; Stefanos Theocharis; Kostas I Lampropoulos; Georgios Kollias; Pantelis Karaiskos
Journal:  J Appl Clin Med Phys       Date:  2022-03-14       Impact factor: 2.243

Review 4.  Radiation Dosimetry by Use of Radiosensitive Hydrogels and Polymers: Mechanisms, State-of-the-Art and Perspective from 3D to 4D.

Authors:  Yves De Deene
Journal:  Gels       Date:  2022-09-19

5.  A novel and effective method for validation and measurement of output factors for Leksell Gamma Knife® Icon™ using TRS 483 protocol.

Authors:  Swapna Lilly Cyriac; Jian Liu; Emel Calugaru; Jenghwa Chang
Journal:  J Appl Clin Med Phys       Date:  2020-09-06       Impact factor: 2.102

6.  Characterization and commissioning of a Leksell Gamma Knife ICON system for framed and frameless stereotactic radiosurgery.

Authors:  Yue-Houng Hu; Susannah V Hickling; Jing Qian; C Robert Blackwell; Luke B McLemore; Erik J Tryggestad
Journal:  J Appl Clin Med Phys       Date:  2022-01-22       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.