Daniel W Bailey1, Lalith Kumaraswamy, Matthew B Podgorsak.
Abstract
Portal dosimetric images acquired for IMRT pretreatment verification show dose errors of up to 15% near the detector edges as compared to dose predictions calculated by a treatment planning system for these off-axis regions. A method is proposed to account for these off-axis effects by precisely correcting the off-axis output factors, which calibrate the imager for absolute dose. Using this method, agreement between the predicted and the measured doses improves by up to 15% for fields near the detector edges, resulting in passing rate improvements of as much as 60% for gamma evaluation of 3 mm, 3% within the collimator jaws.Mesh:
Year: 2009 PMID: 19810481 DOI: 10.1118/1.3187785
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071