Scintillator avalanche photoconductor with high resolution emitter readout for low dose x-ray imaging: lag.

Dan Lia1, Wei Zhao, Masakazu Nanba, Norifomi Egami.   

Abstract

PURPOSE: A new concept of indirect conversion flat-panel imager with avalanche gain and field emitter array (FEA) readout is being investigated. It is referred to as scintillator avalanche photoconductor with high resolution emitter readout (SAPHIRE). The present work investigates the temporal performance, i.e., lag, of SAPHIRE.
METHODS: Since the temporal performance of the x-ray detection materials, i.e., the structured scintillator and avalanche amorphous selenium (a-Se) photoconductor, has been studied previously, the investigation is focused on lag due to the FEA readout method. The principle of FEA readout is similar to that of scanning electron beam readout used in camera tubes, where the dominant source of lag is the energy spread of electrons. Since the principles of emission and beam focusing methods for FEA are different from thermionic emission used in camera tubes, its electron beam energy spread and hence lag is expected to be different. In the present work, the energy spread of the electrons emitted from a FEA was investigated theoretically by analyzing different contributing factors due to the FEA design and operations: The inherent energy spread of field emission, the FEA driving pulse delay, and the angular distribution of emitted electrons. The electron energy spread determined the beam acceptance characteristic curve of the photoconductive target, i.e., the accepted beam current (I(a)) as a function of target potential (V(t)), from which lag could be calculated numerically. Lag calculation was performed using FEA parameters of two prototype HARP-FEA image sensors, and the results were compared with experimental measurements. Strategies for reducing lag in SAPHIRE were proposed and analyzed.
RESULTS: The theoretical analysis shows that the dominant factor for lag is the angular distribution of electrons emitted from the FEA. The first frame lags for two prototype sensors with 4 and 25 microm HARP layer thicknesses were 62.1% and 9.1%, respectively. A lag clearance procedure can be implemented by turning on all the FEA pixels simultaneously between subsequent frames without negative impact of readout speed. For large-area SAPHIRE, the bias electrode for the HARP needs to be divided into strips to allow parallel readout. With typical cardiac detector parameters, SAPHIRE with 128 parallel strips can provide real-time readout (30 frames/s) with first frame lag of -4%.
CONCLUSIONS: The investigation of lag in SAPHIRE shows that the angular distribution of emitted electrons from FEA can result in substantial lag if the readout was performed pixel by pixel. Effective strategies for reducing lag include dividing the bias electrode into multiple strips to allow parallel readout and the incorporation of rapid charge clearance procedure between subsequent frames or rows.

Entities:  

Mesh:

Year:  2009        PMID: 19810477      PMCID: PMC2738744          DOI: 10.1118/1.3187227

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager.

Authors:  J H Siewerdsen; D A Jaffray
Journal:  Med Phys       Date:  1999-08       Impact factor: 4.071

2.  Performance of a 41 x 41 cm2 amorphous silicon flat panel x-ray detector designed for angiographic and R&F imaging applications.

Authors:  Paul R Granfors; Richard Aufrichtig; George E Possin; Brian W Giambattista; Zhong S Huang; Jianqiang Liu; Bing Ma
Journal:  Med Phys       Date:  2003-10       Impact factor: 4.071

3.  Evaluation of the imaging properties of an amorphous selenium-based flat panel detector for digital fluoroscopy.

Authors:  D C Hunt; O Tousignant; J A Rowlands
Journal:  Med Phys       Date:  2004-05       Impact factor: 4.071

4.  Temporal performance of amorphous selenium mammography detectors.

Authors:  Bo Zhao; Wei Zhao
Journal:  Med Phys       Date:  2005-01       Impact factor: 4.071

5.  Feasibility of a large area x-ray sensitive vidicon for medical fluoroscopy: resolution and lag factors.

Authors:  R Luhta; J A Rowlands
Journal:  Med Phys       Date:  1997-05       Impact factor: 4.071

6.  X-ray imaging using amorphous selenium: feasibility of a flat panel self-scanned detector for digital radiology.

Authors:  W Zhao; J A Rowlands
Journal:  Med Phys       Date:  1995-10       Impact factor: 4.071

7.  Fluoroscopic imaging: quantitation of image lag or smearing.

Authors:  J E Gray; J G Stears; F Lopez; M A Wondrow
Journal:  Radiology       Date:  1984-02       Impact factor: 11.105

8.  Indirect flat-panel detector with avalanche gain: fundamental feasibility investigation for SHARP-AMFPI (scintillator HARP active matrix flat panel imager).

Authors:  Wei Zhao; Dan Li; Alla Reznik; B J M Lui; D C Hunt; J A Rowlands; Yuji Ohkawa; Kenkichi Tanioka
Journal:  Med Phys       Date:  2005-09       Impact factor: 4.071

9.  SAPHIRE (scintillator avalanche photoconductor with high resolution emitter readout) for low dose x-ray imaging: spatial resolution.

Authors:  Dan Li; Wei Zhao
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.