The influence of different IMRT techniques on the peripheral dose: a comparison between sMLM-IMRT and helical tomotherapy.

Tilo Wiezorek1, Andrea Schwahofer, Kai Schubert.   

Abstract

PURPOSE: To investigate how segmented multileaf modulation-(sMLM-)based intensity-modulated radiotherapy (IMRT) and dynamic helical tomotherapy (ToTh) affect the peripheral dose (PD) outside the treated region.
MATERIAL AND METHODS: A cuboid Perspex phantom was scanned in a computed tomograph. Different artificial cases were contoured consisting of OARs surrounded by cylindrically shaped planning target volumes (PTVs) with different dimensions. Radiotherapy plans were generated with the sMLM system Konrad (Siemens) and with the ToTh planning system. The plans were optimized in such a way that the dose-volume histograms showed comparable results. The sMLM plans were applied with a linac Primus (Siemens OCS), the ToTh plans with the HiArt system (TomoTherapy); both with 6 MV. Measurements of PDs were performed along the longitudinal axis of the phantom outside the primary beam at different distances from the edge of the PTV (horizontal PD) and also at different depths at a fixed distance from the isocenter (vertical PD). Additional experiments to separate the scatter dose caused by the phantom were performed. This was realized by removing the part of the phantom lying in the primary beam, then applying the same plans like before.
RESULTS: All PD values were normalized to the median dose of the PTV. The PD values for the different PTVs decrease with decreasing PTV size. They also decrease with increasing distance from the isocenter. The horizontal values are in a range of 7% for the largest PTV (diameter = 15 cm) near the primary dose region to 0.2% for the smallest PTV (diameter = 5 cm) far from the primary dose region. The ToTh values are higher than the sMLM values by a maximal factor of 2 near the primary dose region. They become more similar with increasing distance from the edge of the PTV in longitudinal direction. The PD values are nearly equal at a distance of 25 cm from the edge of the PTV. The vertical PDs are higher for the ToTh at depths of > 1 cm but higher for sMLM close to the surface. By removing the scatter cube, the horizontal PD values at middle distances are reduced to one third of the PD values with scatter cube for ToTh (0.5%) and to one half for sMLM (0.8%). This means that without scatter cube the PD for ToTh is lower than that for sMLM. The measured PD values without scatter cube are in the same dimension as published data.
CONCLUSION: The increasing PDs and their trend with increasing PTV size can be explained by Compton scattering of photons from the irradiated volume toward the off-axis measuring points. The further increase of the PD in case of ToTh relative to sMLM is not easy to explain. Different presumptions are possible. The larger field length (in longitudinal direction) of the ToTh plans (consisting of the "real" field length and the overlap) relative to the sMLM plans could be one reason for the higher PD values. The softer energy spectrum of the HiArt machine with more sideward Compton scattering contributions could be another reason.

Entities:  

Mesh:

Year:  2009        PMID: 19806336     DOI: 10.1007/s00066-009-2005-9

Source DB:  PubMed          Journal:  Strahlenther Onkol        ISSN: 0179-7158            Impact factor:   3.621


  22 in total

1.  Evaluation of calculation algorithms implemented in different commercial planning systems on an anthropomorphic breast phantom using film dosimetry.

Authors:  Martin Polednik; Yasser Abo Madyan; Frank Schneider; Dirk Wolff; Burkhardt Bannach; Ulrike Lambrecht; André Wallin; Marian Cwiekala; Klaus Maurer; Florian Reif; Frank Lohr; Frederik Wenz
Journal:  Strahlenther Onkol       Date:  2007-12       Impact factor: 3.621

Review 2.  Accelerated partial breast irradiation: consensus statement of 3 German Oncology societies.

Authors:  Rolf Sauer; Marie-Luise Sautter-Bihl; Wilfried Budach; Petra Feyer; Wolfgang Harms; Rainer Souchan; Diethelm Wollwiener; Rolf Kreienberg; Frederik Wenz
Journal:  Cancer       Date:  2007-09-15       Impact factor: 6.860

3.  The Christie Hospital breast conservation trial: an update at 8 years from inception.

Authors:  G G Ribeiro; B Magee; R Swindell; M Harris; S S Banerjee
Journal:  Clin Oncol (R Coll Radiol)       Date:  1993       Impact factor: 4.126

Review 4.  Accelerated partial breast irradiation in early-stage breast cancer.

Authors:  Mary Ella Sanders; Troy Scroggins; Federico L Ampil; Benjamin D Li
Journal:  J Clin Oncol       Date:  2007-03-10       Impact factor: 44.544

5.  Comparison of wedge versus segmented techniques in whole breast irradiation: effects on dose exposure outside the treatment volume.

Authors:  Veronika Ludwig; Franz Schwab; Matthias Guckenberger; Thomas Krieger; Michael Flentje
Journal:  Strahlenther Onkol       Date:  2008-06       Impact factor: 3.621

6.  Postoperative irradiation of left-sided breast cancer patients and cardiac toxicity. Does deep inspiration breath-hold (DIBH) technique protect the heart?

Authors:  Heidi Stranzl; Brigitte Zurl
Journal:  Strahlenther Onkol       Date:  2008-10-01       Impact factor: 3.621

7.  Accelerated partial-breast irradiation with interstitial implants: the clinical relevance of the calculation of skin doses.

Authors:  Oliver J Ott; Michael Lotter; Rolf Sauer; Vratislav Strnad
Journal:  Strahlenther Onkol       Date:  2007-08       Impact factor: 3.621

8.  Accelerated partial breast irradiation consensus statement from the American Society for Radiation Oncology (ASTRO).

Authors:  Benjamin D Smith; Douglas W Arthur; Thomas A Buchholz; Bruce G Haffty; Carol A Hahn; Patricia H Hardenbergh; Thomas B Julian; Lawrence B Marks; Dorin A Todor; Frank A Vicini; Timothy J Whelan; Julia White; Jennifer Y Wo; Jay R Harris
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-07-15       Impact factor: 7.038

9.  Wide tangential fields including the internal mammary lymph nodes in patients with left-sided breast cancer. Influence of respiratory-controlled radiotherapy (4D-CT) on cardiac exposure.

Authors:  Heidi Stranzl; Brigitte Zurl; Tanja Langsenlehner; Karin S Kapp
Journal:  Strahlenther Onkol       Date:  2009-03-28       Impact factor: 3.621

10.  Improving dose homogeneity in large breasts by IMRT: efficacy and dosimetric accuracy of different techniques.

Authors:  Yasser Abo-Madyan; Martin Polednik; Angelika Rahn; Frank Schneider; Barbara Dobler; Frederik Wenz; Frank Lohr
Journal:  Strahlenther Onkol       Date:  2008-02       Impact factor: 3.621

View more
  19 in total

1.  Dosimetric integration of daily mega-voltage cone-beam CT for image-guided intensity-modulated radiotherapy.

Authors:  A Zabel-du Bois; S Nill; S Ulrich; U Oelfke; B Rhein; P Haering; S Milker-Zabel; A Schwahofer
Journal:  Strahlenther Onkol       Date:  2012-01-05       Impact factor: 3.621

2.  Lack of Relevant Haemogram Changes During Percutaneous Radiotherapy of Localised Prostate Cancer.

Authors:  Felix Zwicker; Benedict Swartman; Peter E Huber; Klaus Herfarth; Jürgen Debus; Henrik Hauswald
Journal:  In Vivo       Date:  2020 May-Jun       Impact factor: 2.155

3.  Second malignancies in high‑dose areas of previous tumor radiotherapy.

Authors:  Birgitta Welte; Peter Suhr; Dirk Bottke; Detlef Bartkowiak; Wolfgang Dörr; Klaus Rüdiger Trott; Thomas Wiegel
Journal:  Strahlenther Onkol       Date:  2010-02-22       Impact factor: 3.621

4.  Treatment results of radiation therapy for muscle-invasive bladder cancer.

Authors:  Tanja Langsenlehner; Carmen Döller; Franz Quehenberger; Heidi Stranzl-Lawatsch; Uwe Langsenlehner; Karl Pummer; Karin S Kapp
Journal:  Strahlenther Onkol       Date:  2010-03-26       Impact factor: 3.621

5.  The impact of direct aperture optimization on plan quality and efficiency in complex head and neck IMRT.

Authors:  Marcello Sabatino; Matthias Kretschmer; Klemens Zink; Florian Würschmidt
Journal:  Radiat Oncol       Date:  2012-01-23       Impact factor: 3.481

6.  Volumetric modulated arc therapy for advanced pancreatic cancer.

Authors:  Wietse Eppinga; Frank Lagerwaard; Wilko Verbakel; Ben Slotman; Suresh Senan
Journal:  Strahlenther Onkol       Date:  2010-06-24       Impact factor: 3.621

7.  CT-myelography for high-dose irradiation of spinal and paraspinal tumors with helical tomotherapy: revival of an old tool.

Authors:  Matthias Uhl; Florian Sterzing; Gregor Habl; Kai Schubert; Gabriele Sroka-Perez; Jürgen Debus; Klaus Herfarth
Journal:  Strahlenther Onkol       Date:  2011-06-27       Impact factor: 3.621

8.  Superiority of conventional intensity-modulated radiotherapy over helical tomotherapy in locally advanced non-small cell lung cancer. A comparative plan analysis.

Authors:  C Song; H Pyo; J Kim; Y K Lim; W C Kim; H J Kim; D W Kim; K H Cho
Journal:  Strahlenther Onkol       Date:  2012-08-17       Impact factor: 3.621

9.  Stereotactic body radiation therapy (SBRT) for treatment of adrenal gland metastases from non-small cell lung cancer.

Authors:  Richard Holy; Marc Piroth; Michael Pinkawa; Michael J Eble
Journal:  Strahlenther Onkol       Date:  2011-03-21       Impact factor: 3.621

10.  Node-positive left-sided breast cancer: does VMAT improve treatment plan quality with respect to IMRT?

Authors:  M Pasler; D Georg; S Bartelt; J Lutterbach
Journal:  Strahlenther Onkol       Date:  2013-03-24       Impact factor: 3.621

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.