Local flip angle correction for improved volume T1-quantification in three-dimensional dGEMRIC using the Look-Locker technique.

Carl Siversson1, Carl-Johan Tiderius, Leif Dahlberg, Jonas Svensson.   

Abstract

PURPOSE: To present an evaluation method for three-dimensional Look-Locker (3D-LL) based T1 quantification, calculating correct T1 values independent of local flip angle (FA) variations. The method was evaluated both in phantoms and in vivo in a delayed Gadolinium Enhanced MRI of Cartilage (dGEMRIC) study with 33 subjects.
MATERIALS AND METHODS: T1 was measured with 3D-LL, using both local FA correction and a precalculated FA slice profile, and compared with standard constant FA correction, for all slices in phantoms and in both femur condyles in vivo. T1 measured using two-dimensional Inversion Recovery (2D-IR) was used as gold standard.
RESULTS: Due to the FA being slice dependent, the standard constant FA correction results in erroneous T1 (systematic error = 109.1 ms in vivo), especially in the outer slices. With local FA correction, the calculated T1 is excellent for all slices in phantoms (<5% deviation from 2D-IR). In vivo the performance is lower (systematic error = -57.5 ms), probably due to imperfect inversion. With precalculated FA correction the performance is very good also in vivo (systematic error = 13.3 ms).
CONCLUSION: With the precalculated FA correction method, the 3D-LL sequence is robust enough for in vivo dGEMRIC, even outside the centermost slices. (c) 2009 Wiley-Liss, Inc.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19787729     DOI: 10.1002/jmri.21906

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  6 in total

1.  Flip angle profile correction for T₁ and T₂ quantification with look-locker inversion recovery 2D steady-state free precession imaging.

Authors:  Mitchell A Cooper; Thanh D Nguyen; Pascal Spincemaille; Martin R Prince; Jonathan W Weinsaft; Yi Wang
Journal:  Magn Reson Med       Date:  2012-01-31       Impact factor: 4.668

Review 2.  Current knowledge and importance of dGEMRIC techniques in diagnosis of hip joint diseases.

Authors:  Christoph Zilkens; Carl Johann Tiderius; Rüdiger Krauspe; Bernd Bittersohl
Journal:  Skeletal Radiol       Date:  2015-04-26       Impact factor: 2.199

3.  Fast B1 Mapping Based on Double-Angle Method with T1 Correction Using Standard Pulse Sequence.

Authors:  Yoshiyuki Ishimori; Takeshi Shimanuki; Tomoya Kobayashi; Masahiko Monma
Journal:  J Med Phys       Date:  2022-03-31

4.  In vivo transport of Gd-DTPA2- into human meniscus and cartilage assessed with delayed gadolinium-enhanced MRI of cartilage (dGEMRIC).

Authors:  Ulf Sigurdsson; Carl Siversson; Eveliina Lammentausta; Jonas Svensson; Carl-Johan Tiderius; Leif E Dahlberg
Journal:  BMC Musculoskelet Disord       Date:  2014-07-09       Impact factor: 2.362

5.  Delayed gadolinium-enhanced MRI of meniscus (dGEMRIM) and cartilage (dGEMRIC) in healthy knees and in knees with different stages of meniscus pathology.

Authors:  Ulf Sigurdsson; Gunilla Müller; Carl Siversson; Eveliina Lammentausta; Jonas Svensson; Carl-Johan Tiderius; Leif E Dahlberg
Journal:  BMC Musculoskelet Disord       Date:  2016-09-29       Impact factor: 2.362

6.  Knee dGEMRIC at 7 T: comparison against 1.5 T and evaluation of T1-mapping methods.

Authors:  Pernilla Peterson; Carl Johan Tiderius; Emma Olsson; Björn Lundin; Lars E Olsson; Jonas Svensson
Journal:  BMC Musculoskelet Disord       Date:  2018-05-16       Impact factor: 2.362

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.