Bilal Yilmaz1, Sakir Arslan, Vedat Akba.
Abstract
A sensitive and efficient method was developed for determination of metoprolol in human plasma by gas chromatography-mass spectrometry (GC-MS). Metoprolol and atenolol (internal standard, IS) were extracted from human plasma with a mixture of ethylacetate and diethylether at basic pH with liquid-liquid extraction. Calibration curves were linear over the concentration range 15-500 ng/ml. Intra- and inter-day precision values for metoprolol in human plasma were less than 6.4, and accuracy (relative error) was better than 8.8%. The analytical recovery of metoprolol from human plasma averaged 91.20%. The limits of detection (LOD) and quantification (LOQ) of metoprolol were 5.0 and 15 ng/ml, respectively. Also the developed and validated GC-MS method was successfully applied to three patients with hypertension who had been given an oral tablet of 100mg metoprolol.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19782235 DOI: 10.1016/j.talanta.2009.06.079
Source DB: PubMed Journal: Talanta ISSN: 0039-9140 Impact factor: 6.057