Literature DB >> 19762248

Study of the IsoAid ADVANTAGE (125)I brachytherapy source dosimetric parameters using Monte Carlo simulation.

Mahdi Sadeghi1, S Hamed Hosseini.   

Abstract

Monte Carlo calculations have been performed using the MCNP4C code for an iodine seed design. As the ADVANTAGE I-125, Model IAI-125 source is commercially available for interstitial brachytherapy treatment, dosimetric characteristics (dose rate constant, radial dose function and anisotropy function) of this source were theoretically determined following the updated AAPM task group 43 (TG-43U1) recommendations. The dose distribution around the seed was calculated with Monte Carlo simulation in liquid water. The Monte Carlo calculated dose rate constant of this source in water was found to be 0.986cGyh(-1)U(-1), with an approximate uncertainty of 0.4%. The obtained result has been compared with the previous study. Comparison of the calculated dose rate constant with the value presented by Meigooni et al. shows a very good agreement. Also the anisotropy function and the radial dose function for this source are graphically compared.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19762248     DOI: 10.1016/j.apradiso.2009.08.007

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

1.  A dosimetry method for low dose rate brachytherapy by EGS5 combined with regression to reflect source strength shortage.

Authors:  Kenichi Tanaka; Kunihiko Tateoka; Osamu Asanuma; Ken-Ichi Kamo; Kaori Sato; Hiromitsu Takeda; Masaru Takagi; Masato Hareyama; Jun Takada
Journal:  J Radiat Res       Date:  2014-01-20       Impact factor: 2.724

2.  Dosimetric Evaluation and Clinical Application of Radioactive Iodine-125 Brachytherapy Stent in the Treatment of Malignant Esophageal Obstruction.

Authors:  Zhe Ji; Qianqian Yuan; Lei Lin; Chao Xing; Xusheng Zhang; Sen Yang; Yuliang Jiang; Haitao Sun; Kaixian Zhang; Junjie Wang
Journal:  Front Oncol       Date:  2022-03-24       Impact factor: 6.244

3.  Percutaneous vertebroplasty combined with interstitial implantation of 125I seeds in banna mini-pigs.

Authors:  Zuozhang Yang; Yu Zhang; Da Xu; Giulio Maccauro; Barbara Rossi; Hua Jiang; Jiaping Wang; Hongpu Sun; Lei Xu; Yanjin Chen; Xuefeng Liu
Journal:  World J Surg Oncol       Date:  2013-02-26       Impact factor: 2.754

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.