Concept, design and pre-clinical studies for remote control percutaneous coronary interventions.

Rafael Beyar1, Tal Wenderow, Doron Lindner, Ganesh Kumar, Rona Shofti.   

Abstract

INTRODUCTION: Percutaneous Coronary Interventions (PCI) is conducted manually with the operator standing at the bed side, exposed to continuous X-ray radiation. A system where the operator can remotely navigate the wire and device during PCI may improve operator safety and convenience and can possibly enhance procedural precision. AIM: To develop a remote navigation system (RNS) that will allow computer controlled, remote manipulation of percutaneous coronary interventions.. METHODS AND
RESULTS: The remote navigation system (RNS) is designed to handle both coronary guide wire and balloon / stent delivery system and can be loaded either with the coronary wire or with both wire and device in parallel. The RNS is comprised of a bedside unit and a remote manipulation unit. The bed side unit has individual wire and device manipulators, capable of precise maneuvering and positioning of the wires and devices. The system was tested in a wire glass model and was evaluated in a normal coronary sheep model. The wire glass model experiments showed that the wire could be navigated to the required branch and the stent/ balloon adequately positioned, as required. The animal experiments showed that the wire could access any required coronary branch and the stent could be adequately positioned under x-ray fluoroscopy, without causing dissection or other vessel trauma.
CONCLUSION: A system that enables remote control percutaneous coronary procedures was developed and tested in vitro and in vivo models. Preclinical studies have demonstrated the feasibility of the concept of remote control PCI and have provided the basis for the pilot clinical study of remote control, stent-assisted PCI.

Entities:  

Year:  2005        PMID: 19758927

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  4 in total

1.  Magnetic resonance imaging compatible remote catheter navigation system with 3 degrees of freedom.

Authors:  M A Tavallaei; M K Lavdas; D Gelman; M Drangova
Journal:  Int J Comput Assist Radiol Surg       Date:  2015-12-24       Impact factor: 2.924

Review 2.  Robotic-assisted angioplasty: current status and future possibilities.

Authors:  Nathaniel R Smilowitz; Giora Weisz
Journal:  Curr Cardiol Rep       Date:  2012-10       Impact factor: 2.931

3.  Statistical Learning Algorithm for in situ and invasive breast carcinoma segmentation.

Authors:  Jagadeesan Jayender; Eva Gombos; Sona Chikarmane; Donnette Dabydeen; Ferenc A Jolesz; Kirby G Vosburgh
Journal:  Comput Med Imaging Graph       Date:  2013-05-19       Impact factor: 4.790

Review 4.  A Review of Robotic Interventional Neuroradiology.

Authors:  C B Beaman; N Kaneko; P M Meyers; S Tateshima
Journal:  AJNR Am J Neuroradiol       Date:  2021-02-04       Impact factor: 4.966

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.