The mechanisms of failure of totally implantable central venous access system: analysis of 73 cases with fracture of catheter.

C H Lin1, H S Wu, D C Chan, C B Hsieh, M H Huang, J C Yu.   

Abstract

BACKGROUND: Totally implantable access ports are often used for the administration of chemotherapy or prolonged intravenous infusions in patients with cancer. The technique has been well described. However, some complications would happen. The pinch-off-syndrome is one of these complications. We report another presentation of pinch-off-syndrome and how to prevent.
METHODS: From January 2005 to December 2007, 73 patients of catheter fracture were collected. The duration of Port-A implantation ranged from January 2003 to October 2007. During this period, 3358 port-catheters were implanted. There were three brands of Port-A implanted included 46% BardPort (Bard, Salt Lake City, UT, USA), 42% A Port (Arrow international, Reading, PA, USA) and 12% PORT-A CATH (Deltec, St. Paul, MN, USA).
RESULTS: The most common clinical presentation was difficulty in injection in 32 cases (43.8%). The incidence of brand C was far lower than brand A and B. The most common site of fracture was at the proximal part (anastomosis between injection port and catheter) in 68 cases (93.2%). The incidence of fracture of Port-A was 6 in 738 (0.81%) in cut-down method; 67 in 2620 (2.56%) in percutaneous subclavian method. Most of the cases (34%) were no more than six months.
CONCLUSION: The most frequent location of fracture Port-A was in proximal part - anastomosis between injection port and catheter. The cause of easily fracture may be associated with pinch-off-syndrome and design of Port-A. This kind of fracture could be prevented by cut-down method and fixed one stitch in proximal part. Copyright (c) 2009 Elsevier Ltd. All rights reserved.

Entities:  

Mesh:

Year:  2009        PMID: 19709847     DOI: 10.1016/j.ejso.2009.07.011

Source DB:  PubMed          Journal:  Eur J Surg Oncol        ISSN: 0748-7983            Impact factor:   4.424


  19 in total

1.  "Difficult to flush chemoport: an important clinical sign".

Authors:  Deepak Sundriyal; Sushil Jain; Suraj Manjunath
Journal:  Indian J Surg Oncol       Date:  2014-09-04

Review 2.  Totally implantable catheter migration and its percutaneous retrieval: case report and review of the literature.

Authors:  E Intagliata; F Basile; R Vecchio
Journal:  G Chir       Date:  2017 Sep-Oct

3.  Ultrasound-guided infraclavicular axillary vein puncture is effective to avoid pinch-off syndrome: a long-term follow-up study.

Authors:  Hideki Osawa; Junichi Hasegawa; Kazuma Yamakawa; Nobuki Matsunami; Shoki Mikata; Junzo Shimizu; Yong Kook Kim; Hirotaka Morishima; Masaki Hirota; Yoshihito Souma; Ho Min Kim; Genta Sawada; Riichiro Nezu
Journal:  Surg Today       Date:  2012-08-28       Impact factor: 2.549

4.  Fracture and migration of implantable venous access port catheters: Cause analysis and management of 4 cases.

Authors:  Shu-Ping Xiao; Bin Xiong; Jun Chu; Xiao-Fang Li; Qi Yao; Chuan-Sheng Zheng
Journal:  J Huazhong Univ Sci Technolog Med Sci       Date:  2015-10-22

5.  Percutaneous retrieval of intravascular venous foreign bodies in children.

Authors:  Anne Marie Cahill; Deddeh Ballah; Paula Hernandez; Lucia Fontalvo
Journal:  Pediatr Radiol       Date:  2011-12-17

6.  Long-term outcomes of totally implantable venous access devices.

Authors:  Yi-Chia Wang; Pei-Lin Lin; Wei-Han Chou; Chih-Peng Lin; Chi-Hsiang Huang
Journal:  Support Care Cancer       Date:  2017-02-09       Impact factor: 3.603

7.  Complicated vascular access port removals: incidence, antecedents and avoidance.

Authors:  Warwick J Teague; Dina Fouad; Fraser D Munro; Amanda J McCabe
Journal:  Pediatr Surg Int       Date:  2015-07-30       Impact factor: 1.827

8.  Embolization of a fractured central venous catheter placed using the internal jugular approach.

Authors:  Atsushi Shimizu; Alan Lefor; Manabu Nakata; Umehachi Mitsuhashi; Masahiro Tanaka; Yoshikazu Yasuda
Journal:  Int J Surg Case Rep       Date:  2014-03-20

9.  Port catheter fracture and migration in Internal Jugular Vein.

Authors:  Rudra Prasad Doley; Preetinder Brar; Sanchit Chaudhary; Rajeev Bedi; Adarsh Chander Swami; Jai Dev Wig
Journal:  Am J Case Rep       Date:  2012-01-19

10.  The Treatment Results of a Standard Algorithm for Choosing the Best Entry Vessel for Intravenous Port Implantation.

Authors:  Wen-Cheng Wei; Ching-Yang Wu; Ching-Feng Wu; Jui-Ying Fu; Ta-Wei Su; Sheng-Yueh Yu; Tsung-Chi Kao; Po-Jen Ko
Journal:  Medicine (Baltimore)       Date:  2015-08       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.