Validation of a hospital-laboratory workstation for immunohematologic methods.

Helge Schoenfeld1, Karin J Pretzel, Christian von Heymann, Bruno Neuner, Ulrich Kalus, Holger Kiesewetter, Axel Pruss.   

Abstract

BACKGROUND: The FREELYS Nano system (Diagast) is a manual workstation for ABO/D grouping, Rh phenotyping, K typing, and antibody screening (ABS) for immunoglobulin G (IgG) antibodies only and works with the erythrocyte-magnetized technology (EMT). The principle of EMT is based on magnetization of red blood cells and avoids centrifugation and washing steps. STUDY DESIGN AND METHODS: A total of 304 samples were tested with our routine blood bank methods, 100 samples for ABO/D grouping, 196 samples (100 at first evaluation, 96 at second evaluation) for Rh phenotyping and K typing (PK7200, Olympus), and 108 samples for ABS (DiaMed). All samples were tested in parallel with the FREELYS Nano.
RESULTS: We found a 100% concordance between the observed (FREELYS Nano) and the expected (Olympus PK7200) results for ABO/D grouping in all 100 samples. For Rh phenotyping and K tests, in 24 of 100 samples false-positive reactions were observed in the first evaluation by the FREELYS Nano. After changing the test kit batch for Rh phenotyping by the manufacturer, a complete concordance in Rh phenotyping and K tests was observed in a second evaluation. For ABS, the FREELYS Nano showed in 4 of 108 samples (3.7%) false-negative reactions for IgG antibodies (two anti-K, one anti-E, one anti-C(w)), and one (0.9%) false-positive reaction.
CONCLUSIONS: The FREELYS Nano is reliably suited to ABO/D grouping, Rh phenotyping, and K testing. The rate of false-negative reactions for IgG antibodies should be reduced.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19709392     DOI: 10.1111/j.1537-2995.2009.02359.x

Source DB:  PubMed          Journal:  Transfusion        ISSN: 0041-1132            Impact factor:   3.157


  5 in total

1.  Performance Evaluation of Automated Immunohematology Analyzer IH-500 for Blood Bank Testing.

Authors:  Sang Hyuk Park; Jaewook Kim; Ji-Hun Lim; Joseph Jeong; Seon-Ho Lee
Journal:  Indian J Hematol Blood Transfus       Date:  2019-05-02       Impact factor: 0.900

2.  Automation in Blood Centre: Its impact on Blood Safety.

Authors:  Snehalata C Gupte
Journal:  Asian J Transfus Sci       Date:  2015-04

Review 3.  Blood Group Testing.

Authors:  Hong-Yang Li; Kai Guo
Journal:  Front Med (Lausanne)       Date:  2022-02-11

4.  Automation in immunohematology.

Authors:  Meenu Bajpai; Ravneet Kaur; Ekta Gupta
Journal:  Asian J Transfus Sci       Date:  2012-07

5.  Comparison Between Conventional and Automated Techniques for Blood Grouping and Crossmatching: Experience from a Tertiary Care Centre.

Authors:  Swarupa Nikhil Bhagwat; Jayashree H Sharma; Julie Jose; Charusmita J Modi
Journal:  J Lab Physicians       Date:  2015 Jul-Dec
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.