Literature DB >> 19700535

Evaluation of automatic exposure control in a direct digital intraoral system.

D Benchimol1, K Näsström, Xq Shi.   

Abstract

OBJECTIVES: The aim of this study was to evaluate the performance of the automatic exposure control (AEC) in an intraoral digital system.
METHODS: Three series of radiographs were captured with the activated AEC on seven dry mandibles adhered to different thicknesses of Plexiglas((R)), which represented soft tissue. Exposure times required for each radiograph of the specimens were analysed as functions of the added Plexiglas thickness. Differences in mean grey levels between pairs of the three images obtained from the same specimen were analysed by employing the subtraction technique. Additionally, four radiographs of all seven mandibles were manually exposed with one layer of Plexiglas, which employed one and two steps below and above the registered automatic exposure time. Six observers evaluated the series of radiographs exposed with one layer of Plexiglas in a random order. The observers were instructed to classify the quality of the diagnostic images as unacceptable, acceptable or excellent.
RESULTS: The results demonstrated a good correlation between the exposure times determined by AEC and Plexiglas thickness (r = 0.85). Mean grey level values for subtraction images obtained from radiographs exposed with different thickness of Plexiglas with activated AEC indicated that almost identical radiographs were obtained, regardless of the Plexiglas thickness. The subjective observer evaluation of the most satisfactory radiographic image quality as a function of exposure time demonstrated that the highest scores were for radiographs exposed with the AEC function activated.
CONCLUSIONS: The AEC function may accurately determine the appropriate exposure times related to object thickness, providing radiographs with satisfactory image quality, and may thus be advantageous in clinical radiographic work.

Mesh:

Year:  2009        PMID: 19700535     DOI: 10.1259/dmfr/22255081

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  3 in total

1.  Auto exposure control (AEC) modality in CBCT devices: Revisiting the reported parameters in current orthodontic literature.

Authors:  Arash Poorsattar Bejeh Mir
Journal:  J Oral Biol Craniofac Res       Date:  2015-08-01

2.  Alternative X-ray filters for an intra-oral digital radiographic system.

Authors:  J Stecke; A D Cruz; S M Almeida; F N Bóscolo
Journal:  Dentomaxillofac Radiol       Date:  2012-01-26       Impact factor: 2.419

3.  Comparison of changes in dental and bone radiographic densities in the presence of different soft-tissue simulators using pixel intensity and digital subtraction analyses.

Authors:  R S de Molon; R G Batitucci; R Spin-Neto; G M Paquier; C E Sakakura; G M Tosoni; G Scaf
Journal:  Dentomaxillofac Radiol       Date:  2013-09-04       Impact factor: 2.419

  3 in total