Primary stability of a hybrid self-tapping implant compared to a cylindrical non-self-tapping implant with respect to drilling protocols in an ex vivo model.

Takeshi Toyoshima1, Wilfried Wagner, Marcus Oliver Klein, Elmar Stender, Marco Wieland, Bilal Al-Nawas.   

Abstract

BACKGROUND: Modifications of implant design have been intending to improve primary stability. However, little is known about investigation of a hybrid self-tapping implant on primary stability. PURPOSES: The aims of this study were to evaluate the primary stability of two hybrid self-tapping implants compared to one cylindrical non-self-tapping implant, and to elucidate the relevance of drilling protocols on primary stability in an ex vivo model.
MATERIALS AND METHODS: Two types of hybrid self-tapping implants (Straumann® Bone Level implant [BL], Straumann® Tapered Effect implant [TE]) and one type of cylindrical non-self-tapping implant (Straumann® Standard Plus implant [SP]) were investigated in the study. In porcine iliac cancellous bones, 10 implants each were inserted either using standard drilling or under-dimensioned drilling protocol. The evaluation of implant-bone interface stability was carried out by records of maximum insertion torque, the Periotest® (Siemens, Bensheim, Germany), the resonance frequency analysis (RFA), and the push-out test.
RESULTS: In each drilling group, the maximum insertion torque values of BL and TE were significantly higher than SP (p=.014 and p=.047, respectively). In each group, the Periotest values of TE were significantly lower than SP (p=.036 and p=.033, respectively). The Periotest values of BL and TE were significantly lower in the group of under-dimensioned drilling than standard drilling (p=.002 and p=.02, respectively). In the RFA, no statistical significances were found in implants between two groups and between implants in each group. In each group, the push-out values of BL and TE were significantly higher than SP (p=.006 and p=.049, respectively).
CONCLUSION: Hybrid self-tapping implants could achieve a high primary stability which predicts them for use in low-density bone. However, there is still a debate to clarify the influence of under-dimensioned drilling on primary stability.
© 2009, Copyright the Authors. Journal Compilation © 2011, Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 19681929     DOI: 10.1111/j.1708-8208.2009.00185.x

Source DB:  PubMed          Journal:  Clin Implant Dent Relat Res        ISSN: 1523-0899            Impact factor:   3.932


  7 in total

1.  Template-guided vs. non-guided drilling in site preparation of dental implants.

Authors:  Uta Scherer; Marcus Stoetzer; Martin Ruecker; Nils-Claudius Gellrich; Constantin von See
Journal:  Clin Oral Investig       Date:  2014-10-30       Impact factor: 3.573

2.  A stepwise under-prepared osteotomy technique improves primary stability in shallow-placed implants: a preliminary study for simultaneous vertical ridge augmentation.

Authors:  Daisuke Ueno; Kei Nakamura; Kousuke Kojima; Takeshi Toyoshima; Hideaki Tanaka; Kazuhiko Ueda; Kiyoshi Koyano; Toshiro Kodama
Journal:  Odontology       Date:  2017-08-02       Impact factor: 2.634

3.  Early outcome of an implant system with a resorbable adhesive calcium-phosphate coating--a prospective clinical study in partially dentate patients.

Authors:  Victor Palarie; Constantin Bicer; Karl M Lehmann; Mohammad Zahalka; Florian G Draenert; Peer W Kämmerer
Journal:  Clin Oral Investig       Date:  2011-08-06       Impact factor: 3.573

4.  Evaluation of sawbones training protocol in bone quality classification using tactile sensation.

Authors:  Tong-Mei Wang; Yu-Chun Lin; Yi-Hao Lan; Li-Deh Lin
Journal:  J Dent Sci       Date:  2022-01-03       Impact factor: 3.719

5.  Effect of implant design on primary stability using torque-time curves in artificial bone.

Authors:  Yoko Yamaguchi; Makoto Shiota; Motohiro Munakata; Shohei Kasugai; Masahiko Ozeki
Journal:  Int J Implant Dent       Date:  2015-08-08

6.  Relevant Design Aspects to Improve the Stability of Titanium Dental Implants.

Authors:  M Herrero-Climent; P López-Jarana; B F Lemos; F J Gil; C Falcão; J V Ríos-Santos; B Ríos-Carrasco
Journal:  Materials (Basel)       Date:  2020-04-18       Impact factor: 3.623

7.  Preoperative evaluation of bone quality for dental implantation using an ultrasound axial transmission device in an ex vivo model.

Authors:  Shinsuke Okada; Akira Kawano; Hiroshi Oue; Yosuke Takeda; Miyuki Yokoi; Katsunori Koretake; Kazuhiro Tsuga
Journal:  Clin Exp Dent Res       Date:  2017-06-09
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.