In vivo dosimetry with a linear MOSFET array to evaluate the urethra dose during permanent implant brachytherapy using iodine-125.

Esther J Bloemen-van Gurp1, Björk K C Haanstra, Lars H P Murrer, Francis C J M van Gils, Andre L A J Dekker, Ben J Mijnheer, Philippe Lambin.   

Abstract

PURPOSE: To develop a technique to monitor the dose rate in the urethra during permanent implant brachytherapy using a linear MOSFET array, with sufficient accuracy and without significantly extending the implantation time. METHODS AND MATERIALS: Phantom measurements were performed to determine the optimal conditions for clinical measurements. In vivo measurements were performed in 5 patients during the (125)I brachytherapy implant procedure. To evaluate if the urethra dose obtained in the operating room with the ultrasound transducer in the rectum and the patient in treatment position is a reference for the total accumulated dose; additional measurements were performed after the implantation procedure, in the recovery room.
RESULTS: In vivo measurements during and after the implantation procedure agree very well, illustrating that the ultrasound transducer in the rectum and patient positioning do not influence the measured dose in the urethra. In vivo dose values obtained during the implantation are therefore representative for the total accumulated dose in the urethra. In 5 patients, the dose rates during and after the implantation were below the maximum dose rate of the urethra, using the planned seed distribution.
CONCLUSION: In vivo dosimetry during the implantation, using a MOSFET array, is a feasible technique to evaluate the dose in the urethra during the implantation of (125)I seeds for prostate brachytherapy.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19665317     DOI: 10.1016/j.ijrobp.2009.04.042

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  On the use of a single-fiber multipoint plastic scintillation detector for 192Ir high-dose-rate brachytherapy.

Authors:  François Therriault-Proulx; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2013-06       Impact factor: 4.071

Review 2.  In vivo dosimetry: trends and prospects for brachytherapy.

Authors:  G Kertzscher; A Rosenfeld; S Beddar; K Tanderup; J E Cygler
Journal:  Br J Radiol       Date:  2014-07-08       Impact factor: 3.039

3.  In vivo real-time rectal wall dosimetry for prostate radiotherapy.

Authors:  Nicholas Hardcastle; Dean L Cutajar; Peter E Metcalfe; Michael L F Lerch; Vladimir L Perevertaylo; Wolfgang A Tomé; Anatoly B Rosenfeld
Journal:  Phys Med Biol       Date:  2010-07-07       Impact factor: 3.609

4.  A Millimeter-scale Single Charged Particle Dosimeter for Cancer Radiotherapy.

Authors:  Kyoungtae Lee; Jessica Scholey; Eric B Norman; Inder K Daftari; Kavita K Mishra; Bruce A Faddegon; Michel M Maharbiz; Mekhail Anwar
Journal:  IEEE J Solid-State Circuits       Date:  2020-09-23       Impact factor: 5.013

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.