Is the Ex-PRESS glaucoma shunt magnetic resonance imaging safe?

Noa Geffen1, Graham E Trope, Tariq Alasbali, David Salonen, Adrian P Crowley, Yvonne M Buys.   

Abstract

INTRODUCTION: Ex-PRESS mini glaucoma shunt is a miniature, stainless steel, nonvalved device used for the treatment of glaucoma. The purpose of this study was to evaluate the safety of the Ex-PRESS glaucoma shunt during magnetic resonance (MR) imaging.
METHODS: The effect of a magnetic field on the Ex-PRESS glaucoma shunt was evaluated in 4 scenarios. (1) Wet dish test: the device was placed floating in a dish filled with water and exposed to magnetic fields of 1.5 and 3 Tesla for 30 seconds. (2) Dry dish test: the device was placed on a dry dish and exposed to a magnetic field of 3 Tesla MR for 30 seconds. (3) Intracameral test: the device was floating freely in the anterior chamber of a human cadaver eye and was exposed to a magnetic field of 3 Tesla MR. (4) Scleral fixation test: the device was implanted through the sclera of a human cadaver eye and was exposed to a magnetic field of 3 Tesla MR.
RESULTS: During the wet dish test the Ex-PRESS glaucoma shunt immediately moved across the dish with both 1.5 and 3 Tesla MR. No movements were noted during the dry dish test, intracameral test, or the scleral fixation test.
CONCLUSIONS: Although the Ex-PRESS glaucoma shunt is influenced by magnetic field forces, it is likely MR of up to 3 Tesla is safe due to resistance provided by the ocular tissue.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 19661826     DOI: 10.1097/IJG.0b013e3181a98bda

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  4 in total

Review 1.  Imaging characteristics of the postoperative globe: a pictorial essay.

Authors:  Yusuke Ito; Ikuo Yamazaki; Yoichi Kikuchi; Eri O'uchi; Toshihiro O'uchi; Hiroaki Kato; Kazuki Hotta
Journal:  Jpn J Radiol       Date:  2016-10-13       Impact factor: 2.374

2.  Ex-PRESS glaucoma filter: an MRI compatible metallic orbital foreign body imaged at 1.5 and 3T.

Authors:  M C Mabray; A Uzelac; J F Talbott; S C Lin; A D Gean
Journal:  Clin Radiol       Date:  2015-02-28       Impact factor: 2.350

3.  Surgical outcomes of the Ex-PRESS glaucoma filtration device in African American and white glaucoma patients.

Authors:  Sarwat Salim; Haiming Du; Sumalee Boonyaleephan; Jim Wan
Journal:  Clin Ophthalmol       Date:  2012-06-22

Review 4.  EX-PRESS Glaucoma Filtration Device: efficacy, safety, and predictability.

Authors:  Jessica E Chan; Peter A Netland
Journal:  Med Devices (Auckl)       Date:  2015-09-02
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.