| Literature DB >> 19649494 |
Maria Carolina Braga Norte Esteves1, Ana Helena Bannwart Dell' Aringa, Gustavo Viani Arruda, Alfredo Rafael Dell' Aringa, José Carlos Nardi.
Abstract
UNLABELLED: Brainstem Evoked Response Audiometry (BERA) is an objective and non-invasive method of hearing assessment which detects electrical activity from the inner ear to the inferior colliculus. AIM: To assess the hearing pathway in normal hearing individuals and compare differences associated with gender, age and ear side (left and right). STUDYEntities:
Mesh:
Year: 2009 PMID: 19649494 PMCID: PMC9445975
Source DB: PubMed Journal: Braz J Otorhinolaryngol ISSN: 1808-8686
Mean and standard deviation (ms) of absolute latencies (n = 120 ears).
| Wave | Wave | Wave | Interval | Interval | Interval | |
|---|---|---|---|---|---|---|
| I | III | V | I - III | III - V | I - V | |
| Mean | 1,69 | 3,82 | 5,59 | 2,13 | 1,78 | 3,90 |
| Standard | ± | ± | ± | ± 0,14 | ± 0,18 | ± 0,21 |
| deviation | 0,13 | 0,16 | 0,20 |
Chart 1Comparison between sexes according to the ear (n = 120 ears) - * = p < 0.05
Mean, standard deviation (ms) and p value according to sex (n= 120 ears).
| Male RE | Fem RE | Value p | Male LE | Fem LE | p Value | |
|---|---|---|---|---|---|---|
| Wave I | 1.66± 0,13 | 1.69± 0,13 | 0.841 | 1.7± 0,16 | 1.7± 0,12 | 0.80 |
| Wave III | 3.86± 0,16 | 3.79± 0,15 | 0.992 | 3.82± 0,16 | 3.83± 0,16 | 0.996 |
| Wave V | 5.67± 0,25 | 5.53± 0,17 | 0.047 | 5.66± 0,21 | 5.58± 0,18 | 0.537 |
| Wave I-III | 2.18± 0,15 | 2.1± 0,16 | 0.723 | 2.12± 0,13 | 2.13± 0,18 | 0.580 |
| Wave III-V | 1.81± 0,19 | 1.75± 0,17 | 0.447 | 1.84± 0,15 | 1.75± 0,15 | 0.937 |
| Onda I-V | 4.01± 0,29 | 3.85± 0,15 | 0.007 | 3.96± 0,22 | 3.97± 0,19 | 0.571 |
Right ear means (ms) in females and p value (n= 39 ears).
| Females | I | III | V | I-III | III-V | I-V |
|---|---|---|---|---|---|---|
| RE < 35 a | 1,72 | 3,80 | 5,53 | 2,08 | 1,73 | 3,80 |
| RE > 35 a | 1,69 | 3,85 | 5,61 | 2,15 | 1,76 | 3,92 |
| p Value | 0,30 | 0,36 | 0,12 | 0,17 | 0,38 | 0,00 |
Chart 2Comparison between the means found and means suggested by the manual of the device (n = 60 patients) - * = p < 0.05
Difference among absolute latency means (ms), the p value and the confidence interval (CI) (n = 60 patients)
| Wave I | Wave III | Wave V | Interval I - V | |
|---|---|---|---|---|
| Difference among means | 0,10 | 0,18 | 0,02 | 0,08 |
| p value | 0,000 | 0,000 | 0,209 | 0,000 |
| CI (95%) | 0,08 – 0,13 | 0,15 – 0,21 | -0,01 – 0,06 | -0,11 - −0,04 |
Chart 3Comparison between the means found and means suggested by Fukuda et al.'s study (n = 120 ears) - * = p < 0.05
Difference among absolute latency means (ms), the p value and the confidence interval (CI) in the right ear (RE) (n= 60 ears).
| RE | I | III | V | I - III | III - V | I - V |
|---|---|---|---|---|---|---|
| Difference between means | 0,07 | 0,05 | 0,01 | 0,02 | 0,05 | 0,05 |
| p value | 0,000 | 0,005 | 0,147 | 0,316 | 0,102 | 0,056 |
| CI (95%) | -0,12 a −0,05 | -0,10 a −0,20 | -0,10 a 0,01 | -0,02 a 0,06 | -0,01 a 0,08 | 0,00 a 0,11 |
Difference among absolute latency means (ms), the p value and the confidence interval (CI) in the left ear (LE) (n= 60 ears).
| LE | I | III | V | I - III | III - V | I - V |
|---|---|---|---|---|---|---|
| Difference between means | 0,07 | 0,04 | 0,06 | 0,01 | 0,00 | 0,02 |
| p value | 0,009 | 0,141 | 0,194 | 0,759 | 0,655 | 0,631 |
| CI (95%) | -0,08 a −0,01 | -0,07 a 0,01 | -0,08 a 0,02 | -0,04 a 0,05 | -0,03 a 0,05 | -0,04 a 0,06 |