Literature DB >> 19629728

A comparison of vital signs charted by nurses with automated acquired values using waveform quality indices.

Monica Sapo1, Shaozhi Wu, Shadnaz Asgari, Norma McNair, Farzad Buxey, Neil Martin, Xiao Hu.   

Abstract

OBJECTIVE: (1) To investigate if there exist any discrepancies between the values of vital signs charted by nurses and those recorded by bedside monitors for a group of patients admitted for neurocritical care. (2) To investigate possible interpretations of discrepancies by exploring information in the alarm messages and the raw waveform data from monitors.
METHODS: Each charted vital sign value was paired with a corresponding value from data collected by an archival program of bedside monitors such that the automatically archived data preceded the charted data and had minimal time lag to the charted value. Next, the absolute differences between the paired values were taken as the discrepancy between charted and automatically-collected data. Archived alarm messages were searched for technical alarms of sensor/lead failure types. Additionally, 7-min waveform data around the place of large discrepancy were analyzed using signal abnormality indices (SAI) for quantifying the quality of recorded signals.
RESULTS: About 31,145 pairs of systolic blood pressure (BP-S) and 67,097 pairs of SpO(2) were investigated. Seven and a half percent of systolic blood pressure pairs had a discrepancy greater than 20 mmHg and less than one percent of the SpO2 pairs had a discrepancy greater than 10. We could not find any technical alarms from the monitors that could explain the large difference. However, SAI calculated for the waveforms associated with this group of cases was significantly larger than the SAI values calculated for the control waveform data of the same patients with small discrepancies.
CONCLUSION: Charted vital signs reflect in large the raw data as reported by bedside monitors. Poor signal quality could partially explain the existence of cases of large discrepancies.

Entities:  

Mesh:

Year:  2009        PMID: 19629728     DOI: 10.1007/s10877-009-9192-x

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  17 in total

1.  Arterial blood pressure and heart rate discrepancies between handwritten and computerized anesthesia records.

Authors:  D L Reich; R K Wood; R Mattar; M Krol; D C Adams; S Hossain; C A Bodian
Journal:  Anesth Analg       Date:  2000-09       Impact factor: 5.108

Review 2.  Digital electronic communication between ICU ventilators and computers and printers.

Authors:  T D East; W H Young; R M Gardner
Journal:  Respir Care       Date:  1992-09       Impact factor: 2.258

3.  Computers in the intensive care unit.

Authors:  Mitchell M Levy
Journal:  J Crit Care       Date:  2004-12       Impact factor: 3.425

4.  Assessing data quality in manual entry of ventilator settings.

Authors:  David K Vawdrey; Reed M Gardner; R Scott Evans; James F Orme; Terry P Clemmer; Loren Greenway; Frank A Drews
Journal:  J Am Med Inform Assoc       Date:  2007-02-28       Impact factor: 4.497

5.  The impact of a clinical information system in an intensive care unit.

Authors:  Abele Donati; Vincenzo Gabbanelli; Simona Pantanetti; Paola Carletti; Tiziana Principi; Benedetto Marini; Simonetta Nataloni; Gisella Sambo; Paolo Pelaia
Journal:  J Clin Monit Comput       Date:  2007-11-24       Impact factor: 2.502

6.  A comparison of nursing activities associated with manual and automated documentation in an Australian intensive care unit.

Authors:  C Marasovic; C Kenney; D Elliott; D Sindhusake
Journal:  Comput Nurs       Date:  1997 Jul-Aug

7.  A robust approach toward recognizing valid arterial-blood-pressure pulses.

Authors:  Shadnaz Asgari; Marvin Bergsneider; Xiao Hu
Journal:  IEEE Trans Inf Technol Biomed       Date:  2009-10-30

Review 8.  The accuracy of clinical information systems.

Authors:  Nicholas S Ward
Journal:  J Crit Care       Date:  2004-12       Impact factor: 3.425

9.  An algorithm for extracting intracranial pressure latency relative to electrocardiogram R wave.

Authors:  Xiao Hu; Peng Xu; Darrin J Lee; Paul Vespa; Kevin Baldwin; Marvin Bergsneider
Journal:  Physiol Meas       Date:  2008-03-17       Impact factor: 2.833

10.  Morphological clustering and analysis of continuous intracranial pressure.

Authors:  Xiao Hu; Peng Xu; Fabien Scalzo; Paul Vespa; Marvin Bergsneider
Journal:  IEEE Trans Biomed Eng       Date:  2008-11-07       Impact factor: 4.538

View more
  5 in total

1.  Multimodal noninvasive monitoring of soft tissue wound healing.

Authors:  Michael Bodo; Timothy Settle; Joseph Royal; Eric Lombardini; Evelyn Sawyer; Stephen W Rothwell
Journal:  J Clin Monit Comput       Date:  2013-07-06       Impact factor: 2.502

2.  Acute vital signs changes are underrepresented by a conventional electronic health record when compared with automatically acquired data in a single-center tertiary pediatric cardiac intensive care unit.

Authors:  Adam W Lowry; Craig A Futterman; Avihu Z Gazit
Journal:  J Am Med Inform Assoc       Date:  2022-06-14       Impact factor: 7.942

3.  A subspace decomposition approach toward recognizing valid pulsatile signals.

Authors:  Shadnaz Asgari; Peng Xu; Marvin Bergsneider; Xiao Hu
Journal:  Physiol Meas       Date:  2009-10-01       Impact factor: 2.833

4.  Smoothing Effect in Vital Sign Recordings: Fact or Fiction? A Retrospective Cohort Analysis of Manual and Continuous Vital Sign Measurements to Assess Data Smoothing in Postoperative Care.

Authors:  Hamish R Tomlinson; Marco A F Pimentel; Stephen Gerry; David A Clifton; Lionel Tarassenko; Peter J Watkinson
Journal:  Anesth Analg       Date:  2018-10       Impact factor: 5.108

5.  Big Data for Clinical Trials: Automated Collection of SpO2 for a Trial of Oxygen Targets during Mechanical Ventilation.

Authors:  Kevin G Buell; Jonathan D Casey; Li Wang; Jonathan P Wanderer; Wesley H Self; Todd W Rice; Matthew W Semler
Journal:  J Med Syst       Date:  2020-07-31       Impact factor: 4.460

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.