Literature DB >> 19622898

Accuracy of sphygmomanometers at pharmacies.

Dilek Gurgenyatagi Erdem1, Emre Erdem, Melda Dilek, Turkan Aydogdu, Nevzat Selim, Feride Demiray, Aysel Sari, Tekin Akpolat.   

Abstract

BACKGROUND/AIMS: The aims of this study are to assess the reasons of using sphygmomanometers at pharmacies and to evaluate their accuracy.
METHODS: 135 devices (118 aneroid, 1 mercury, and 16 automated) from 125 pharmacies (located in Samsun city center) were included in the study. A non-randomized, cross-sectional design was used for the study protocol which had two parts: assessment of devices and a questionnaire about the pharmacy and present sphygmomanometer(s).
RESULTS: 40 (30%) of the 135 sphygmomanometers were inaccurate. 65 (48%) of the devices were older than 1 year and there was no correlation between the duration of the ownership of the sphygmomanometers and their inaccuracy (p > 0.05). Blood pressure measurement is a frequent practice at pharmacies. The aneroid type of sphygmomanometers was common. A limited number of devices were checked for accuracy before. The number of validated devices was low.
CONCLUSION: Training programs for pharmacists including the accuracy of sphygmomanometers and regular checks of sphygmomanometers for accuracy will be beneficial to the community and to the subjects requesting measurement of blood pressure at the pharmacies. (c) 2009 S. Karger AG, Basel.

Entities:  

Mesh:

Year:  2009        PMID: 19622898     DOI: 10.1159/000228936

Source DB:  PubMed          Journal:  Kidney Blood Press Res        ISSN: 1420-4096            Impact factor:   2.687


  3 in total

Review 1.  Physician-pharmacist collaborative practice and telehealth may transform hypertension management.

Authors:  Stefano Omboni; Mauro Tenti; Claudio Coronetti
Journal:  J Hum Hypertens       Date:  2018-12-13       Impact factor: 3.012

2.  Lancet Commission on Hypertension group position statement on the global improvement of accuracy standards for devices that measure blood pressure.

Authors:  James E Sharman; Eoin O'Brien; Bruce Alpert; Aletta E Schutte; Christian Delles; Michael Hecht Olsen; Roland Asmar; Neil Atkins; Eduardo Barbosa; David Calhoun; Norm R C Campbell; John Chalmers; Ivor Benjamin; Garry Jennings; Stéphane Laurent; Pierre Boutouyrie; Patricio Lopez-Jaramillo; Richard J McManus; Anastasia S Mihailidou; Pedro Ordunez; Raj Padwal; Paolo Palatini; Gianfranco Parati; Neil Poulter; Michael K Rakotz; Clive Rosendorff; Francesca Saladini; Angelo Scuteri; Weimar Sebba Barroso; Myeong-Chan Cho; Ki-Chul Sung; Raymond R Townsend; Ji-Guang Wang; Tine Willum Hansen; Gregory Wozniak; George Stergiou
Journal:  J Hypertens       Date:  2020-01       Impact factor: 4.776

3.  [Lancet Commission on Hypertension Group position statement on the global improvement of accuracy standards for devices that measure blood pressurePosicionamento do Grupo da Lancet Commission on Hypertension sobre a melhoria global dos padrões de acurácia para aparelhos que medem a pressão arterial].

Authors:  James E Sharman; Eoin O'Brien; Bruce Alpert; Aletta E Schutte; Christian Delles; Michael Hecht Olsen; Roland Asmar; Neil Atkins; Eduardo Barbosa; David Calhoun; Norm R C Campbell; John Chalmers; Ivor Benjamin; Garry Jennings; Stéphane Laurent; Pierre Boutouyrie; Patricio Lopez-Jaramillo; Richard J McManus; Anastasia S Mihailidou; Pedro Ordunez; Raj Padwal; Paolo Palatini; Gianfranco Parati; Neil Poulter; Michael K Rakotz; Clive Rosendorff; Francesca Saladini; Angelo Scuteri; Weimar Sebba Barroso; Myeong-Chan Cho; Ki-Chul Sung; Raymond R Townsend; Ji-Guang Wang; Tine Willum Hansen; Gregory Wozniak; George Stergiou
Journal:  Rev Panam Salud Publica       Date:  2020-03-01
  3 in total