Literature DB >> 19622852

Real-time profiling of respiratory motion: baseline drift, frequency variation and fundamental pattern change.

D Ruan1, J A Fessler, J M Balter, P J Keall.   

Abstract

To precisely ablate tumor in radiation therapy, it is important to locate the tumor position in real time during treatment. However, respiration-induced tumor motions are difficult to track. They are semi-periodic and exhibit variations in baseline, frequency and fundamental pattern (oscillatory amplitude and shape). In this study, we try to decompose the above-mentioned components from discrete observations in real time. Baseline drift, frequency (equivalently phase) variation and fundamental pattern change characterize different aspects of respiratory motion and have distinctive clinical indications. Furthermore, smoothness is a valid assumption for each one of these components in their own spaces, and facilitates effective extrapolation for the purpose of estimation and prediction. We call this process 'profiling' to reflect the integration of information extraction, decomposition, processing and recovery. The proposed method has three major ingredients: (1) real-time baseline and phase estimation based on elliptical shape tracking in augmented state space and Poincaré sectioning principle; (2) estimation of the fundamental pattern by unwarping the observation with phase estimate from the previous step; (3) filtering of individual components and assembly in the original temporal-displacement signal space. We tested the proposed method with both simulated and clinical data. For the purpose of prediction, the results are comparable to what one would expect from a human operator. The proposed approach is fully unsupervised and data driven, making it ideal for applications requiring economy, efficiency and flexibility.

Entities:  

Mesh:

Year:  2009        PMID: 19622852     DOI: 10.1088/0031-9155/54/15/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  19 in total

1.  High-quality t2-weighted 4-dimensional magnetic resonance imaging for radiation therapy applications.

Authors:  Dongsu Du; Shelton D Caruthers; Carri Glide-Hurst; Daniel A Low; H Harold Li; Sasa Mutic; Yanle Hu
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-03-30       Impact factor: 7.038

2.  Effect of secondary particles on image quality of dynamic flat panels in carbon ion scanning beam treatment.

Authors:  S Mori; S Amano; T Furukawa; T Shirai; K Noda
Journal:  Br J Radiol       Date:  2014-12-23       Impact factor: 3.039

3.  Objective function to obtain multiple representative waveforms for a novel helical CT scan protocol.

Authors:  Dan Ruan; David Thomas; Daniel A Low
Journal:  Med Phys       Date:  2015-03       Impact factor: 4.071

4.  Motion management strategies and technical issues associated with stereotactic body radiotherapy of thoracic and upper abdominal tumors: A review from NRG oncology.

Authors:  Edward D Brandner; Indrin J Chetty; Tawfik G Giaddui; Ying Xiao; M Saiful Huq
Journal:  Med Phys       Date:  2017-04-20       Impact factor: 4.071

5.  Respiratory motion prediction and prospective correction for free-breathing arterial spin-labeled perfusion MRI of the kidneys.

Authors:  Hao Song; Dan Ruan; Wenyang Liu; V Andrew Stenger; Rolf Pohmann; Maria A Fernández-Seara; Tejas Nair; Sungkyu Jung; Jingqin Luo; Yuichi Motai; Jingfei Ma; John D Hazle; H Michael Gach
Journal:  Med Phys       Date:  2017-02-21       Impact factor: 4.071

6.  A novel fast helical 4D-CT acquisition technique to generate low-noise sorting artifact-free images at user-selected breathing phases.

Authors:  David Thomas; James Lamb; Benjamin White; Shyam Jani; Sergio Gaudio; Percy Lee; Dan Ruan; Michael McNitt-Gray; Daniel Low
Journal:  Int J Radiat Oncol Biol Phys       Date:  2014-03-07       Impact factor: 7.038

7.  On a PCA-based lung motion model.

Authors:  Ruijiang Li; John H Lewis; Xun Jia; Tianyu Zhao; Weifeng Liu; Sara Wuenschel; James Lamb; Deshan Yang; Daniel A Low; Steve B Jiang
Journal:  Phys Med Biol       Date:  2011-08-24       Impact factor: 3.609

8.  Real-time prediction of tumor motion using a dynamic neural network.

Authors:  Majid Mafi; Saeed Montazeri Moghadam
Journal:  Med Biol Eng Comput       Date:  2020-01-08       Impact factor: 2.602

9.  A Bayesian approach for three-dimensional markerless tumor tracking using kV imaging during lung radiotherapy.

Authors:  Chun-Chien Shieh; Vincent Caillet; Michelle Dunbar; Paul J Keall; Jeremy T Booth; Nicholas Hardcastle; Carol Haddad; Thomas Eade; Ilana Feain
Journal:  Phys Med Biol       Date:  2017-03-21       Impact factor: 3.609

10.  Real time 4D IMRT treatment planning based on a dynamic virtual patient model: proof of concept.

Authors:  Bingqi Guo; X George Xu; Chengyu Shi
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

View more